Your browser doesn't support javascript.
loading
Monitoreo de presión intracraneana: infección y otras complicaciones con el uso de K.30 subdural y fibra óptica intraparenquimatosa / Intracranial pressure monitoring: infection an other complicactions with the subdural K-30 and the intraparenquimatous fiber optic
Gardella, Javier L; Guevara, Martín; Purves, Cynthia; Moughty Cueto, Carolina.
  • Gardella, Javier L; Hospital General de Agudos Juan A. Fernández.
  • Guevara, Martín; Hospital General de Agudos Juan A. Fernández.
  • Purves, Cynthia; Hospital General de Agudos Juan A. Fernández.
  • Moughty Cueto, Carolina; Hospital General de Agudos Juan A. Fernández.
Rev. argent. neurocir ; 20(4): 151-155, oct.-dic. 2006. tab, graf
Artigo em Espanhol | LILACS | ID: lil-451751
RESUMEN
Objetivo. Evaluar las complicaciones en la utilización de trasductores de presión intracraneana (PIC) K-30 subdurales (KS) y fibras ópticas intraparenquimatosas (FO). Método. En un estudio estadístico descriptivo y retrospectivo se revisaron las historias clínicas de los pacientes que ingresaron a Terapia Intensiva del Hospital "Juan A. Fernández" (junio 1993 - febrero 2003) y se requirieron monitoreo de PIC. Los datos fueron procesados con el programa STATA 6.0. Resultados. Se efectuaron 252 monitoreos en 191 pacientes 71 monitores fueron FO y 181 fueron SK; 188 fueron catéteres únicos y 64 recambios. Los motivos de recambio fueron arrancamiento, error de lectura, deterioro neurológico y colocación >5 días. El arrancamiento ocurrió en 13 casos (12 KS y 1 FO); error de lectura o falla técnica en 33 casos (20 KS y 13 FO); fístula de LCR en 9 casos (7 KS y 2FO). Sólo en un caso de los que presentaron fístula se constató cultivo de LCR positivo. (En 6 casos el recambio se debió a deterioro neurológico. Recambio por monitoreo >5 días ocurrió en 12). Período de monitorización 1-15 días. La permanencia del monitor en pacientes con catéteres sin infección fue de 3.62 días y en pacientes con infección 4.67 días. En 29 casos se comprobó infección (10FO y 19KS). No fue significativa la diferencia entre edad, sexo, Glasgow al ingreso y cirugía a cielo abierto cuando de lo relacionó con infección y tipo de monitor. Fue significativa la incidencia de infección en los pacientes reoperados. Conclusión. Técnicamente el monitoreo PIC con KS subdural demostró ser un método confiable. La incidencia de complicaciones, en particular la infecciosa, con sistema KS fue similar a las observadas con FO. Palabras clave catéter subdural, complicaciones, infección, presión intracraniana
ABSTRACT

Objective:

To evaluate the complications between to (sic) types of intracranial pressure monitoring devices subdural K-30 (SK) and intraparenchymal fiber optic (FO).

Method:

In this descriptive retrospective study we reviewed the clinical records of those patients qho were admitted to the Intensive Care Unit of the Hospital "Juan A. Fernandez" (June 1993-February 2003) and requiered intracranial pressure monitoring. We processed the data with the STATA 6.0 programme.

Results:

We inserted 252 monitors in 191 patients 71 FO and 181 SK; 188 were primary an 64 were secondary procedures. The causes of removal were dislocation of the device, technical failure, neurological deteroration and premanence longer than 5 days. Dislocation occurred in 13 cases (12 SK, 1 FO), technical failure was observed in 33 cases (20 SK, 13 FO). We observed cerebrospinal fluid fistula (CSF) in 9 cases (7 SK, 2 FO). Only 1 case that presented with fistula had positive CSF culture with a non typified germen. In 6 cases the change of the catheter was due to neurological deteroration. Changing due to permanence longer than 5 days occurred in 12 cases. The period of monitorization was 1-15 days (median 3 days). The permanence of the catheter in non infected patients was 3.62±3.22 days and on infected patients was 4.67±2.30 days. On 29 cases we observed infection, 10 FO and 19 SK (NS p). It was not significant the difference between age, sex, Glasgow and surgery when related to infection and type of device. It was significant the rate of infection in patients that had to be reoperated.

Conclusion:

From a technical point of view monitorization of the intracranial pressure using SK proved to be reliable. In our experience the rate of complications between the SK was similar to those observed with FO. Key words intracranial pressure - complications - infection - subdural catheter.