Your browser doesn't support javascript.
loading
Controle do tempo de protrombina em sangue capilar e venoso em pacientes com anticoagulação oral: correlação e concordância / Comparative study of a portable system for prothrombin monitoring using capillary blood against venous blood measurements in patients using oral anticoagulants: correlation and concordance
Leiria, Tiago Luiz Luz; Pellanda, Lúcia Campos; Magalhães, Eros; Lima, Gustavo Glotz de.
  • Leiria, Tiago Luiz Luz; Fundação Universitária de Cardiologia. Instituto de Cardiologia do Rio Grande do Sul. Ambulatório de Anticoagulação. Serviço de Eletrofisiologia. Porto Alegre. BR
  • Pellanda, Lúcia Campos; Fundação Universitária de Cardiologia. Instituto de Cardiologia do Rio Grande do Sul. Ambulatório de Anticoagulação. Serviço de Eletrofisiologia. Porto Alegre. BR
  • Magalhães, Eros; Fundação Universitária de Cardiologia. Instituto de Cardiologia do Rio Grande do Sul. Ambulatório de Anticoagulação. Serviço de Eletrofisiologia. Porto Alegre. BR
  • Lima, Gustavo Glotz de; Fundação Universitária de Cardiologia. Instituto de Cardiologia do Rio Grande do Sul. Ambulatório de Anticoagulação. Serviço de Eletrofisiologia. Porto Alegre. BR
Arq. bras. cardiol ; 89(1): 1-5, jul. 2007. graf, tab
Artigo em Português | LILACS | ID: lil-459808
RESUMO
FUNDAMENTO O uso de anticoagulantes orais (ACO) é comum na prática cardiológica, tendo como principais indicações a fibrilação atrial e próteses valvares. Os pacientes em uso de ACO necessitam de controle freqüente do tempo de protrombina (TP). Novos sistemas portáteis de monitorização, para medida do tempo de protrombina, sem necessidade da coleta de sangue por venopunção, facilitam a rotina desses pacientes.

OBJETIVO:

Comparar as medidas do TP realizadas pelo sistema Coaguchek S®, em sangue capilar, com o método padrão em sangue venoso.

MÉTODOS:

Cento e vinte e sete pacientes em acompanhamento no ambulatório de anticoagulação do Instituto de Cardiologia foram submetidos ao método convencional e coleta de sangue capilar da polpa digital para medida com o sistema Coaguchek S®.

RESULTADOS:

A idade foi de 58±14 anos, 90 por cento eram brancos. As indicações de ACO foram fibrilação atrial 49,6 por cento e próteses valvares 37,0 por cento. O coeficiente de correlação r s foi 0,90 (p<0,0001; IC95 por cento 0,87-0,93) entre o sistema em estudo e o controle. A medida de concordância entre pacientes com INR <2, entre 2 e 3,5 e > 3,5 foi de Kappa 73,5 por cento. O sistema Coaguchek S® superestimou o INR em 0,15±0,85 unidades. Valores acima de 3,5 unidades de INR mostram discrepância importante entre as duas técnicas.

CONCLUSÕES:

O sistema Coaguchek S® mostrou boa correlação e bom grau de concordância quando comparado com o controle, com resultados menores que 4. Porém, valores de INR acima de 3,5 ainda necessitam de confirmação com o método padrão.
ABSTRACT

BACKGROUND:

Oral anticoagulants (OAC) are widely used in cardiology and are mainly indicated in cases of atrial fibrillation and prosthetic heart valves. Regular prothrombin time (PT) control is required for patients using OAC. New portable monitoring systems for measuring prothrombin time, eliminate the need to collect blood by venous puncture and facilitate daily life for these patients.

OBJECTIVE:

To compare PT measurements using the Coaguchek S™ system with capillary blood and the standard method in venous blood.

METHODS:

One hundred and twenty-seven patients from the Cardiology Institute's anticoagulation clinic underwent conventional blood collection and capillary blood collection via a finger prick for measurements using the Coaguchek S™ system.

RESULTS:

The mean age was 58 ± 14 years and 90 percent of the patients were white. OAC indications were atrial fibrillation (49.6 percent) and prosthetic heart valves (37.0 percent). The correlation coefficient, r s , was 0.90 (p<0.0001; CI95 percent 0.87-0.93) between the Coaguchek S™ system and the control method. The Kappa measure of agreement among the patients with INR <2, INR between 2 and 3.5 and INR > 3.5 was 73.5 percent. The Coaguchek S™ system overestimated INR by 0.15±0.85 units. A great deal of discrepancy was found between the two techniques for INR values higher than 3.5 units.

CONCLUSION:

The Coaguchek S™ system when compared to the control method revealed good correlation and a high degree of agreement for results lower than 4 units. However, confirmation is required for INR values above 3.5 using the standard method.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Tempo de Protrombina / Coagulação Sanguínea / Monitoramento de Medicamentos / Cardiopatias / Anticoagulantes Tipo de estudo: Estudo observacional / Estudo de prevalência / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Arq. bras. cardiol Assunto da revista: Cardiologia Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Fundação Universitária de Cardiologia/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Tempo de Protrombina / Coagulação Sanguínea / Monitoramento de Medicamentos / Cardiopatias / Anticoagulantes Tipo de estudo: Estudo observacional / Estudo de prevalência / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Arq. bras. cardiol Assunto da revista: Cardiologia Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Fundação Universitária de Cardiologia/BR