Your browser doesn't support javascript.
loading
Influência da metodologia de avaliação da camada semi-redutora em radiologia diagnóstica / The methodology for evaluating half-value layer and its influence on the diagnostic radiology
Lacerda, Marco Aurélio de Sousa; Silva, Teógenes Augusto da; Oliveira, Arno Heeren de.
  • Lacerda, Marco Aurélio de Sousa; Centro de Desenvolvimento da Tecnologia Nuclear. Comissão Nacional de Energia Nuclear. Belo Horizonte. BR
  • Silva, Teógenes Augusto da; Centro de Desenvolvimento da Tecnologia Nuclear. Comissão Nacional de Energia Nuclear. Belo Horizonte. BR
  • Oliveira, Arno Heeren de; Universidade Federal de Minas Gerais. Departamento de Energia Nuclear. Belo Horizonte. BR
Radiol. bras ; 40(5): 331-336, set.-out. 2007. graf, tab
Artigo em Inglês, Português | LILACS | ID: lil-467769
RESUMO

OBJETIVO:

A análise crítica da metodologia de medida da camada semi-redutora em feixes de raios X teve como base o regulamento técnico para proteção radiológica e controle de qualidade em radiodiagnóstico médico e odontológico.

MATERIAIS E MÉTODOS:

Na medida da camada semi-redutora, a técnica radiográfica, o arranjo experimental, os meios espalhadores, a instrumentação, o tamanho de campo de radiação e a metodologia de análise dos dados foram considerados.

RESULTADOS:

A camada semi-redutora obtida em condições de boa geometria, para a técnica escolhida, foi de 2,44 ± 0,02 mmAl. Em relação a este valor, observaram-se desvios máximos na camada semi-redutora de 4,1 por cento na variação da geometria, de 98,8 por cento na variação da câmara de ionização e do meio espalhador, e de até 29,5 por cento com outro tipo de detector.

CONCLUSÃO:

Os resultados mostraram que não ocorre variação significante na camada semi-redutora para diferentes tamanhos de campo de radiação, mas foram evidentes a influência do espalhamento na superestimação da camada semi-redutora e a redução desta na presença de blocos de chumbo como meio espalhador. O procedimento prático adotado mostrou-se bastante confiável e evidenciou a grande discrepância decorrente da adoção de metodologias impróprias, enfatizando a necessidade de estabelecer um procedimento padrão para a medida da camada semi-redutora.
ABSTRACT

OBJECTIVE:

The critical analysis of the methodology for evaluating the x-ray beam half-value layer was based on technical standards for radiological protection and quality control in medical and dental diagnoses. MATERIALS AND

METHODS:

Exposure parameters, experimental set-up, scattering materials, instrumentation, size of radiation field and methodology of data analysis have been taken into consideration for measurements of half-value layer.

RESULTS:

The half-value layer, in good geometry, measured 2.44 ± 0.02 mmAl. A maximum deviation of 4.1 percent was found under other geometry conditions. A 98.8 percent deviation was observed with a different ionization chamber, and scattering material, and 29.5 percent with another type of detector.

CONCLUSION:

Results showed that no relevant variation is observed in the half-value layer for different radiation field sizes. On the other hand, the influence of the radiation scattering on the half-value layer overestimation was clearly observed, as well as the half-value layer reduction in the presence of lead used as scattering material. The practical methodology adopted for half-value layer evaluation has shown to be quite reliable, besides evidencing remarkable discrepancies resulting from the adoption of inappropriate methodologies, emphasizing the necessity of establishing a standard procedure for measuring the half-value layer.
Assuntos
Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Proteção Radiológica / Radiologia / Radiometria / Tecnologia Radiológica Tipo de estudo: Estudo diagnóstico / Guia de Prática Clínica Idioma: Inglês / Português Revista: Radiol. bras Assunto da revista: Diagnóstico por Imagem / Radiologia Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro de Desenvolvimento da Tecnologia Nuclear/BR / Universidade Federal de Minas Gerais/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Proteção Radiológica / Radiologia / Radiometria / Tecnologia Radiológica Tipo de estudo: Estudo diagnóstico / Guia de Prática Clínica Idioma: Inglês / Português Revista: Radiol. bras Assunto da revista: Diagnóstico por Imagem / Radiologia Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro de Desenvolvimento da Tecnologia Nuclear/BR / Universidade Federal de Minas Gerais/BR