Your browser doesn't support javascript.
loading
Uso da monitorização contínua de glicose (CGMS) por quatro (96 horas) ou cinco (120 horas) dias em pacientes com DM1: existe vantagem? / Continuous glucose monitoring system (CGMS) in type 1 diabetic patients during 4 (96h) or 5 (120h) days: there is advantage?
Maia, Frederico F. R; Araújo, Levimar R.
  • Maia, Frederico F. R; Faculdade de Ciências Médicas de Minas Gerais. Departamento de Fisiologia. Belo Horizonte. BR
  • Araújo, Levimar R; FCMMG. Hospital Universitário São José. Clínica de Endocrinologia. Belo Horizonte. BR
Arq. bras. endocrinol. metab ; 52(3): 499-505, abr. 2008. tab
Artigo em Português | LILACS | ID: lil-482579
RESUMO

OBJETIVO:

Avaliar a acurácia, as complicações e o impacto no controle glicêmico de pacientes com diabetes melito tipo 1 (DM1) submetidos ao uso da monitorização contínua de glicose (CGMS) por quatro e cinco dias.

MÉTODOS:

Estudamos 36 pacientes DM1 (44,5 por centoM/55,5 por centoF), divididos em três grupos, com idade, tempo de DM e A1c semelhantes (p < 0,05), submetidos ao CGMS por 72 h (G1), 96 h (G2) e 120 h (G3). Foram analisados GC média e pelo sensor, número de leituras, coeficiente de correlação, mediana da diferença absoluta (MAD por cento); complicações (trauma, infecção, desconexão, abandono); hiperglicemia pós-prandial (HPP) e hipoglicemia assintomática (< 70 mg/dl). Os níveis de A1c após três meses e um ano foram determinados nos três grupos.

RESULTADOS:

Não houve diferença técnica entre os grupos coeficiente de correlação > 0,79 e MAD por cento < 28 por cento (p < 0,01) em 95 por cento. O uso do sensor por > 72 h não se associou com erro de sinal, desconexão, trauma, alarmes ou infecção local (p < 0,01). Verificou-se alta correlação entre GC média e pelo sensor, nos três grupos (p = 0,01). A hipoglicemia silenciosa foi mais duradoura (min.) à noite versus dia (p = 0,05), em todos os grupos. A HPP foi estatisticamente mais detectada nos grupos 1 e 3. Observou-se redução significante da A1c três meses após a CGMS no G1 (72 h) e G3 (120 h) (p < 0,001 e p = 0,002, respectivamente), que se manteve após um ano (p < 0,001 e p = 0,047, respectivamente).

CONCLUSÃO:

Evidenciamos alta acurácia/eficácia técnica com baixo índice de complicações em pacientes submetidos ao CGMS por 96 h ou 120 h. Não verificamos benefícios em relação à CGMS por 72 h quanto à redução da A1c em curto (três meses) e médio (um ano) prazo. O sensor CGMS pode ser utilizado por > 72 h, sem prejuízo técnico, mas sem grandes benefícios do ponto de vista clínico, para pacientes com DM1.
ABSTRACT

BACKGROUND:

To evaluate the accuracy, complications and impact in glycemic control in type 1 diabetic patients (DM1) submitted to 4 or 5 days of CGMS.

METHODS:

We studied 36 DM1 patients (44.5 percentM/55.5 percentF), in three groups without no difference about age, DM duration and A1c levels (p < 0.05), submitted to 72h (G1), 96h (G2) and 120h (G3) CGMS profile. It were analyzed capillary glycemia (CG) and mean CGMS sensor’s glycemic value; correlation coefficient, median absolute percent difference (MAD percent), number of sensor reading, complications (trauma, local infection, disconnection, dropped), postprandial hyperglycemia, unrecognized hypoglycemia (< 70 mg/dl). A1c levels were measured at the start (1 month before) and after 3 and 12 months in each group.

RESULTS:

No technical difference were observed into 3 groups correlation coefficient > 0.79 and MAD < 28 percent in 95 percent (p < 0.01). The use of CGMS sensor more than 72h was not related to signal error, trauma, local infection or disconnection. The mean capillary glucose values showed no difference by glucose CGMS sensor (p = 0.01) in all groups. The nighttime episodes of hypoglycemia lasted longer (min) than daytime episodes in all groups (p = 0.05). The postprandial hyperglycemia was statistically identified in groups 1 and 3. This data showed significant decreased A1c level three months after the CGMS in G1 (72h) and G3 (120h) (p < 0.001 and p = 0.002, respectively), which sustained after 1 year (p < 0.001 e p = 0.047, respectively).

CONCLUSIONS:

The CGMS showed to be a very safety method, with high accuracy/technical efficacy in patients undergoing 96 h and 120 h of CGMS. We do not observed advantages in the use of CGMS during 96 h or 120 h against 72 h in decrease A1c levels after 3 and 12 months. It is possible the use of CGMS > 72h, with no technical damage. However, we do not observed significant clinical benefits of this conduct in DM1 patients.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Glicemia / Automonitorização da Glicemia / Monitorização Ambulatorial / Diabetes Mellitus Tipo 1 Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Adolescente / Adulto / Criança / Criança, pré-escolar / Humanos País/Região como assunto: América do Sul / Brasil Idioma: Português Revista: Arq. bras. endocrinol. metab Assunto da revista: Endocrinologia / Metabolismo Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: FCMMG/BR / Faculdade de Ciências Médicas de Minas Gerais/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Glicemia / Automonitorização da Glicemia / Monitorização Ambulatorial / Diabetes Mellitus Tipo 1 Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Adolescente / Adulto / Criança / Criança, pré-escolar / Humanos País/Região como assunto: América do Sul / Brasil Idioma: Português Revista: Arq. bras. endocrinol. metab Assunto da revista: Endocrinologia / Metabolismo Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: FCMMG/BR / Faculdade de Ciências Médicas de Minas Gerais/BR