In vitro integrity of implant external hexagon after application of surgical placement torque simulating implant locking
Davi, Letícia Resende; Golin, Alexsander Luiz; Bernardes, Sérgio Rocha; Araújo, Cleudmar Amaral de; Neves, Flávio Domingues.
  • Davi, Letícia Resende; s.af
  • Golin, Alexsander Luiz; Pontifical Catholic University of Curitiba. School of Mechanical Engineering. Department of Mechanical Engineering. BR
  • Bernardes, Sérgio Rocha; Latin American Institute of Dental Research and Education. Scientific Department. BR
  • Araújo, Cleudmar Amaral de; Federal University of Uberlândia. School of Mechanical Engineering. Department of Projects and Mechanical Systems. BR
  • Neves, Flávio Domingues; Federal University of Uberlândia. School of Dentistry. Department of Occlusion. BR
Braz. oral res ; 22(2): 125-131, 2008. ilus, graf, tab
Artigo em Inglês | LILACS | ID: lil-485951
ABSTRACT
The aim of this study was to evaluate the integrity of the external hexagon of an implant system with internal and external hexagons but with prosthetic connection through the external hexagon (Internal Torque, IT) in comparison with that of an implant system with external hexagon with mount (External Hexagon, EH). A device was made to measure the rotational freedom angles between implant and abutment hexagons in 10 implants from each group after the application of surgical placement torques of 45, 60 and 80 Ncm simulating implant locking. The distances between the vertices of the external hexagon were also obtained. Rotational freedom data were subjected to ANOVA and Tukey's test (P < .05) showing no significant difference between the angles of the intact implants (EH - 3.31 ± 0.41° and IT - 3.30 ± 0.17°) and after application of a 45 Ncm torque (EH - 3.27 ± 0.38° and IT - 3.31 ± 0.22°). However, after application of a 60 Ncm torque there were significant differences (IT - 3.40 ± 0.20° and EH - 4.03 ± 0.54°). After application of a 80 Ncm torque, the IT implant presented values of 3.39 ± 0.21° whereas the EH did not support the torque, suffering deformation of its external hexagon. Within the limits of this study, it can be concluded that the IT implant system may be preferable in clinical situations where implant placement within a certain bone density could generate torques higher than 60 Ncm.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Implantes Dentários / Dente Suporte / Planejamento de Prótese Dentária / Prótese Dentária Fixada por Implante Idioma: Inglês Revista: Braz. oral res Assunto da revista: Odontologia Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Federal University of Uberlândia/BR / Latin American Institute of Dental Research and Education/BR / Pontifical Catholic University of Curitiba/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Implantes Dentários / Dente Suporte / Planejamento de Prótese Dentária / Prótese Dentária Fixada por Implante Idioma: Inglês Revista: Braz. oral res Assunto da revista: Odontologia Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Federal University of Uberlândia/BR / Latin American Institute of Dental Research and Education/BR / Pontifical Catholic University of Curitiba/BR