Método simplificado para la titulación de la presión del CPAP en pacientes con diagnóstico de Síndrome de Apneas Hipopneas del Sueño (SAHS). Serie de casos empleando el equipo autoajustable Autoset T / Simplified method of CPAP titration in patients with diagnosis of obstructive sleep apnea. Case series using the autoadjustable Autoset T
Rey de Castro Mujica, Jorge; Rosales Mayor, Edmundo; Ferreyra Pereyra, Jenny Rocío.
  • Rey de Castro Mujica, Jorge; Universidad Peruana Cayetano Heredia. Facultad de Medicina Alberto Hurtado. Lima. PE
  • Rosales Mayor, Edmundo; s.af
  • Ferreyra Pereyra, Jenny Rocío; Centro de Trastornos Respiratorios del Sueño. Lima. PE
Rev. méd. hered ; 19(1): 18-24, ene.-mar. 2008. graf
Artigo em Espanhol | LILACS, LIPECS | ID: lil-499701
RESUMEN
El método estándar actualmente aceptado para determinar la presión del CPAP es la titulación convencional por polisomnografía. El procedimiento consume tiempo, requiere de personal calificado y tiene costos operativos elevados.

Objetivos:

Evaluar el uso del CPAP auto ajustable para la titulación de la presión del CPAP en pacientes con diagnóstico del Síndrome de Apneas Hipopneas del Sueño (SAHS) moderado a severo. Material y

métodos:

Se presentan 54 pacientes con diagnóstico de SAHS documentado por polisomnografía (6) y poligrafía respiratoria (48). Para la titulación se empleó el equipo auto ajustable Autoset T (ResMed). Todos los pacientes recibieron en el consultorio las instrucciones para el uso del equipo antes de llevarlo a su domicilio. Si los resultados de la primera prueba no eran satisfactorios, el paciente volvía a llevarlo en una segunda oportunidad.

Resultados:

Se realizaron 54 titulaciones, por cada paciente. En la primera prueba ocho estudios (15 por ciento) fallaron, por lo que estos pacientes llevaron el equipo a su domicilio para una segunda prueba. Finalmente, sólo dos pruebas (3.7 por ciento) la titulación fue frustra y en estos casos se requirió una titulación convencional por polisomnografía. La presión final programada en el CPAP convencional se determinó de acuerdo a la lectura visual de la gráfica. Al momento de cierre del registro, 44 (81 por ciento) de los pacientes titulados con este método usaban regularmente el CPAP.

Conclusiones:

El método no convencional de titulación basado en la lectura visual del registro en el CPAP auto ajustable es útil, sencillo, muy económico y aplicable en nuestro entorno.
ABSTRACT
The accepted Standard method for titration actually accepted is the convencional polysomnography. It consumes time, qualified personnel and has a high operational cost. Objetive To evaluate the auto adjustable CPAP use for CPAP titration in patients with diagnosis of moderate to severe Obstructive Sleep Apnea (OSA). Mterial and

methods:

We report 54 patients with the diagnosis of OSA done by polysomnography (6) and respiratory polygraphy (48). We used the Autoset T (Resmed) autoadjustable system for titration. All patients received during an office interview the instructions about how to use the equipment before they were taken it home. If results of the first test were not satisfactory, the patient would take it home for a second test.

Results:

54 studies were done, one for each patient. In the first test, 8 studies (15 per cent) failed, so they took the equipment home once again. Finally, only in 2 patients (3.7 per cent) the titration failed and they needed a conventional titration with polysomnography. The final pressure to be programmed to the CPAP was defined according to the graphicÆs visual reading. At the moment we finished the data base, 44 (81 per cent) patients were using the CPAP regularly.

Conclusions:

We consider that non-conventional titration based on the visual reading of auto adjustable CPAP recors is an usefull, simple, very cheap CPAP titration method in our country.
Assuntos
Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Síndromes da Apneia do Sono / Respiração com Pressão Positiva / Polissonografia Tipo de estudo: Estudo diagnóstico Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. méd. hered Ano de publicação: 2008 Tipo de documento: Artigo Instituição/País de afiliação: Centro de Trastornos Respiratorios del Sueño/PE / Universidad Peruana Cayetano Heredia/PE

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Síndromes da Apneia do Sono / Respiração com Pressão Positiva / Polissonografia Tipo de estudo: Estudo diagnóstico Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. méd. hered Ano de publicação: 2008 Tipo de documento: Artigo Instituição/País de afiliação: Centro de Trastornos Respiratorios del Sueño/PE / Universidad Peruana Cayetano Heredia/PE