Comparação da resistência imposta pelos filtros artificiaisdurante a ventilação mecânica / Comparison of the resistance imposed by artificial filters during mechanical ventilation
Ribeiro, Denise Cardoso; Lopes, Célia Regina; Lima, Roberta Zancani de; Teixeira, Yara Cordeiro do Nascimento.
  • Ribeiro, Denise Cardoso; Faculdades Metropolitanas Unidas. Curso de Fisioterapia. São Paulo. BR
  • Lopes, Célia Regina; Centro Universitário do Triângulo. Curso de Fisioterapia. Uberlândia. BR
  • Lima, Roberta Zancani de; Faculdades Metropolitanas Unidas. Curso de Fisioterapia. São Paulo. BR
  • Teixeira, Yara Cordeiro do Nascimento; Faculdades Metropolitanas Unidas. Curso de Fisioterapia. São Paulo. BR
Arq. méd. ABC ; 32(Supl. 2): S42-S46, dez. 2007. tab, graf
Artigo em Português | LILACS | ID: lil-508296
RESUMO

Introdução:

Os filtros trocadores de calor e umidificadores (HMEs)higroscópicos, hidrofóbicos e mistos, que exercem as funções das viasaéreas superiores na ventilação mecânica, podem promover aumento daresistência no circuito aumentando o trabalho respiratório do paciente.O objetivo deste estudo foi comparar a resistência imposta entre osfiltros higroscópico (Medsize) e misto (Gibeck) ao sistema respiratórioartificial, durante a assistência ventilatória mecânica.

Método:

Os filtrosda marca Medsize e Gibeck foram testados entre o intermediário Y eo balão de teste, sendo analisada a pressão antes (Pmax) e após cadafiltro, com diferentes fluxos e quantidade de solução salina a 0,9% .

Resultados:

Observou-se resistência em ambos os filtros com aumentode no mínimo 1 cmH2O. O filtro da marca Gibeck apresentou maiorresistência, 10 cmH2O, em relação ao filtro da marca Medsize, 5 cmH2O.

Conclusão:

Os filtros conectados ao circuito ventilatório artificialapresentaram aumento de resistência relacionado com o incremento deinstilação de solução salina a 0,9 e com maiores valores de fluxo.
ABSTRACT

Introduction:

Hygroscopic, hydrophobic and coed heat moistureexchangers (HMEs) that prevent the loss of moisture and heat duringprolonged mechanical ventilation may increase resistance and breathingworkload. The objective of this estudy was to compare the increaseof resistance between hygroscopic (Medsize) and coed filter (Gibeck)through an artificial circuit during mechanical ventilation assistance.

Methods:

Medsize and Gibeck filters were tested between the Yintermediary and the test balloon. The pressure before (Pmax) andafter the filter were analysed with different air flow and increments of 0.9% saline.

Results:

In both filters was observed a resistance increaseof 1 cmH2O. Gibeck filter showed higher resistance, 10 cmH2O, whencompared with Medsize filter, 5 cmH2O.

Conclusion:

This studyshowed that filters resistance were increased with increments of 0.9% saline and higher values was observed with air flows.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Respiração Artificial Idioma: Português Revista: Arq. méd. ABC Assunto da revista: Medicina Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro Universitário do Triângulo/BR / Faculdades Metropolitanas Unidas/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Respiração Artificial Idioma: Português Revista: Arq. méd. ABC Assunto da revista: Medicina Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro Universitário do Triângulo/BR / Faculdades Metropolitanas Unidas/BR