Estudo comparativo da qualidade de imagem dos modos de aquisição da PET: validação de um protocolo para reduzir a dose de radiação / Comparative study on imaging quality in PET acquisition modes: validation of a protocol for reducing the radiation dose
Radiol. bras
;
42(2): 103-107, mar.-abr. 2009. ilus, graf, tab
Artigo
em Português
| LILACS
| ID: lil-513151
RESUMO
OBJETIVO:
O presente experimento visa a validar um protocolo de aquisição em 3D na tomografia por emissão de pósitrons, em substituição ao modo 2D, de forma a reduzir a dose de radiação nos pacientes, sem perda da qualidade de imagens.MATERIAIS E MÉTODOS:
Foram realizadas 27 simulações em equipamento Discovery ST, nos modos 2D com quatro minutos de aquisição e 3D com dois e quatro minutos. Utilizou-se um simulador do protocolo da National Electrical Manufacturers Association. No interior deste simulador estão inseridas seis esferas com diferentes diâmetros para a determinação da qualidade de imagem. As aquisições foram comparadas por três médicos nucleares, sem que eles identificassem o modo de aquisição. Cada observador atribuiu o valor igual a 1 quando alguma esfera não foi identificada ou valor 2 para esferas visíveis.RESULTADOS:
A análise qualitativa pelo kappa generalizado demonstrou que a frequência de esferasvisíveis foi maior no modo 3D com quatro minutos (85%) e a porcentagem de concordância também foi maior (88,9%), com kappa generalizado = 0,725 [0,507;0,942].CONCLUSÃO:
O modo 3D com quatro minutos de aquisição e com menores atividades de FDG-18F pode ser utilizado em pacientes com biótipo equivalente ao simulador, sem perda de qualidade de imagem.ABSTRACT
OBJECTIVE:
The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS ANDMETHODS:
The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizinga chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres withdifferent diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres.RESULTS:
The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942].CONCLUSION:
The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality.
Texto completo:
DisponíveL
Índice:
LILACS (Américas)
Assunto principal:
Imageamento Tridimensional
/
Tomografia por Emissão de Pósitrons
Tipo de estudo:
Estudo prognóstico
/
Pesquisa qualitativa
Idioma:
Português
Revista:
Radiol. bras
Assunto da revista:
Diagnóstico por Imagem
/
Radiologia
Ano de publicação:
2009
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Hospital Albert Einstein/BR
/
Hospital Israelita Albert Einstein/BR
/
Universidade Federal de São Paulo/BR
Similares
MEDLINE
...
LILACS
LIS