Estudo comparativo da qualidade de imagem dos modos de aquisição da PET: validação de um protocolo para reduzir a dose de radiação / Comparative study on imaging quality in PET acquisition modes: validation of a protocol for reducing the radiation dose
Nogueira, Solange Amorim; Lederman, Henrique Manoel; Wagner, Jairo; Yamaga, Lilian Yuri; Cunha, Marcelo Livorsi da; Funari, Marcelo Buarque de Gusmão.
  • Nogueira, Solange Amorim; Hospital Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Lederman, Henrique Manoel; Universidade Federal de São Paulo. Escola Paulista de Medicina. Departamento de Diagnóstico por Imagem. Disciplina de Diagnósticopor Imagem em Pediatria. São Paulo. BR
  • Wagner, Jairo; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Yamaga, Lilian Yuri; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Cunha, Marcelo Livorsi da; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Funari, Marcelo Buarque de Gusmão; Hospital Israelita Albert Einstein. Departamento de Imagem. São Paulo. BR
Radiol. bras ; 42(2): 103-107, mar.-abr. 2009. ilus, graf, tab
Artigo em Português | LILACS | ID: lil-513151
RESUMO

OBJETIVO:

O presente experimento visa a validar um protocolo de aquisição em 3D na tomografia por emissão de pósitrons, em substituição ao modo 2D, de forma a reduzir a dose de radiação nos pacientes, sem perda da qualidade de imagens.

MATERIAIS E MÉTODOS:

Foram realizadas 27 simulações em equipamento Discovery ST, nos modos 2D com quatro minutos de aquisição e 3D com dois e quatro minutos. Utilizou-se um simulador do protocolo da National Electrical Manufacturers Association. No interior deste simulador estão inseridas seis esferas com diferentes diâmetros para a determinação da qualidade de imagem. As aquisições foram comparadas por três médicos nucleares, sem que eles identificassem o modo de aquisição. Cada observador atribuiu o valor igual a 1 quando alguma esfera não foi identificada ou valor 2 para esferas visíveis.

RESULTADOS:

A análise qualitativa pelo kappa generalizado demonstrou que a frequência de esferasvisíveis foi maior no modo 3D com quatro minutos (85%) e a porcentagem de concordância também foi maior (88,9%), com kappa generalizado = 0,725 [0,507;0,942].

CONCLUSÃO:

O modo 3D com quatro minutos de aquisição e com menores atividades de FDG-18F pode ser utilizado em pacientes com biótipo equivalente ao simulador, sem perda de qualidade de imagem.
ABSTRACT

OBJECTIVE:

The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS AND

METHODS:

The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizinga chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres withdifferent diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres.

RESULTS:

The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942].

CONCLUSION:

The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Imageamento Tridimensional / Tomografia por Emissão de Pósitrons Tipo de estudo: Estudo prognóstico / Pesquisa qualitativa Idioma: Português Revista: Radiol. bras Assunto da revista: Diagnóstico por Imagem / Radiologia Ano de publicação: 2009 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Hospital Albert Einstein/BR / Hospital Israelita Albert Einstein/BR / Universidade Federal de São Paulo/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Imageamento Tridimensional / Tomografia por Emissão de Pósitrons Tipo de estudo: Estudo prognóstico / Pesquisa qualitativa Idioma: Português Revista: Radiol. bras Assunto da revista: Diagnóstico por Imagem / Radiologia Ano de publicação: 2009 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Hospital Albert Einstein/BR / Hospital Israelita Albert Einstein/BR / Universidade Federal de São Paulo/BR