Comparison of five portable peak flow meters
Clinics
;
65(5): 469-474, 2010. ilus, tab
Artigo
em Inglês
| LILACS
| ID: lil-548626
ABSTRACT
OBJECTIVE: To compare the measurements of spirometric peak expiratory flow (PEF) from five different PEF meters and to determine if their values are in agreement. Inaccurate equipment may result in incorrect diagnoses of asthma and inappropriate treatments. METHODS: Sixty-eight healthy, sedentary and insufficiently active subjects, aged from 19 to 40 years, performed PEF measurements using Air Zone®, Assess®, Galemed®, Personal Best® and Vitalograph® peak flow meters. The highest value recorded for each subject for each device was compared to the corresponding spirometric values using Friedman's test with Dunn's post-hoc (p<0.05), Spearman's correlation test and Bland-Altman's agreement test. RESULTS: The median and interquartile ranges for the spirometric values and the Air Zone®, Assess®, Galemed®, Personal Best® and Vitalograph® meters were 428 (263-688 L/min), 450 (350-800 L/min), 420 (310-720 L/min), 380 (300-735 L/min), 400 (310-685 L/min) and 415 (335-610 L/min), respectively. Significant differences were found when the spirometric values were compared to those recorded by the Air Zone® (p<0.001) and Galemed ® (p<0.01) meters. There was no agreement between the spirometric values and the five PEF meters. CONCLUSIONS: The results suggest that the values recorded from Galemed® meters may underestimate the actual value, which could lead to unnecessary interventions, and that Air Zone® meters overestimate spirometric values, which could obfuscate the need for intervention. These findings must be taken into account when interpreting both devices' results in younger people. These differences should also be considered when directly comparing values from different types of PEF meters.
Texto completo:
DisponíveL
Índice:
LILACS (Américas)
Assunto principal:
Asma
/
Pico do Fluxo Expiratório
/
Fluxômetros
Tipo de estudo:
Estudos de avaliação
Limite:
Adulto
/
Feminino
/
Humanos
/
Masculino
Idioma:
Inglês
Revista:
Clinics
Assunto da revista:
Medicina
Ano de publicação:
2010
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Centro Universitário Central Paulista/BR
/
Federal University of São Carlos/BR
Similares
MEDLINE
...
LILACS
LIS