Titulación domiciliaria de CPAP en el síndrome de apnea del sueño: ¿una, 3 ó 7 noches? / Home auto-titrating positive airway (APAP) for sleep apnea syndrome: one, three or seven nights?
Gallego, Claudio; Simkin, Pablo; Menéndez, Pablo.
  • Gallego, Claudio; Policlínica Bancaria. Servicio de Neumonología. Ciudad Autónoma de Buenos Aires. AR
  • Simkin, Pablo; Policlínica Bancaria. Ciudad Autónoma de Buenos Aires. AR
  • Menéndez, Pablo; Policlínica Bancaria. Ciudad Autónoma de Buenos AIres. AR
Rev. am. med. respir ; 9(3): 133-139, sept. 2009. tab, graf
Artigo em Espanhol | LILACS | ID: lil-554452
RESUMEN
Los equipos automáticos de CPAP (APAP) permiten establecer la presión eficaz de tratamiento en pacientes con apnea del sueño, sin la necesidad de supervisión y con un menor costo que las titulaciones realizadas bajo control polisomnográfico. Entre estos, el Autoset (ResMed) modifica la presión entregada en base a la presencia de apneas y limitación al flujo inspiratorio. Pacientes con síndrome de apnea/hipopnea del sueño (SAHS) fueron estudiados con este equipo durante 1 semana, luego de un período de al menos 60 días de uso de CPAP con una presión arbitraria entre 8 y 9 cmH2O. Comparamos el valor de presión obtenido en 1, 3 y 7 noches de titulación. Se estableció un cumplimiento mínimo aceptable de 6 noches y un nivel de fuga promedio < 0.4 l/seg. Ingresaron al estudio 24 pacientes de los cuales se excluyeron 2 (uno por fuga elevada, otro por usarlo < 6 noches). Los 22 pacientes estudiados (3 mujeres): edad 61±11 años, IMC 32.2±5 kg/m2, circunferencia de cuello 44.5±4.7cm, IAH 45±20/h (n=18), Indice de desaturación de 4% 36±12/h (n=4). El APAP fue usado por 16 pacientes las 7 noches. El tiempo de uso promedio en 1 vs. 3 vs. 7 noches fue de 7.1±1.2 vs. 7.0±0.9 vs. 6.7±1.1 h/noche (p<0.05 al comparar 1 y 3 vs. 7 noches). La presión eficaz para abolir los eventos obstructivos durante el 95% del trazado (P95) fue de 9.6±2.6 vs. 9.9±2.3 vs. 9.8±2.3 cmH2O (p NS). Analizando en cada paciente las P95 obtenidas en los distintos períodos estudiados, encontramos diferencias menor o - =1 cmH2O en 6 pacientes entre la primera y el promedio de 7 noches (p<0.05), y en 1 paciente comparando 3 y 7 noches (p NS). Conclusión: El registro de 3 noches mostró resultados similares a los obtenidos con 7 noches de autotitulación. Por el contrario, una sola noche resultó insuficiente debido a significativa variabilidad.
ABSTRACT
Automatic devices for auto-titrating positive airway pressure (APAP) enable to estimate the effective pressure level of treatment in patients with sleep apnea, without supervision by a technician and at a lower cost than attended polysomnography titration. One of these devices, Autoset (ResMed), modifies the delivered pressure based on the presence of apneas and inspiratory flow limitation. Patients with sleep apnea-hypopnea syndrome were studied with this device during a week, after more than 60 days of continuous positive airway pressure (CPAP) with anarbitrary pressure between 8 and 9 cm H20. The study compared the pressure level measured in 1, 3 and 7 nights. The recordings were considered acceptable if at least 6 nights were included, and the mean leak was less than 0.4 l/second. Twenty four patients were recruited for the study; 2 were excluded (one by high leak level, and another by using the equipment < 6 nights). The 22 patients included in the analysis were 3 females and 19 males: age 61 ±11 years, BMI 32.2 ±5 kg/m2, neck circumference 44.5 ±4.7cm, Apnea Hypopnea Index (AHI) 45 ±20/h (n=18), oxygen desaturation index of 4% 36 ±12/h (n=4). The APAP was used across 7 nights by 16 patients. Mean time of use in 1 versus 3 versus 7 nights was respectively 7.1 ±1.2, 7.0 ±0.9 and 6.7 ±1.1 h/night (p < 0.05 comparing 1 and 3 versus 7 nights). Effective pressure to abolish obstructive events during 95% of the registered time (P95) was respectively 9.6 ±2.6, 9.9 ±2.3 and 9.8 ±2.3 cm water (pNS). Analyzing the individual P95 in the different periods, there was a difference of - > 1cm H20 in 6 patients between the first night and the mean of 7 nights (p < 0.05); in 1 patient this difference was observed comparing 3 and 7 nights (p NS). Conclusion: The registers of autotitrating during 3 nights and 7 nights were similar. On the other hand the observation of only one night was insufficient because of significant variability.
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Síndromes da Apneia do Sono / Respiração com Pressão Positiva / Pressão Positiva Contínua nas Vias Aéreas Limite: Humanos Idioma: Espanhol Revista: Rev. am. med. respir Assunto da revista: Medicina / Pneumologia Ano de publicação: 2009 Tipo de documento: Artigo País de afiliação: Argentina Instituição/País de afiliação: Policlínica Bancaria/AR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Síndromes da Apneia do Sono / Respiração com Pressão Positiva / Pressão Positiva Contínua nas Vias Aéreas Limite: Humanos Idioma: Espanhol Revista: Rev. am. med. respir Assunto da revista: Medicina / Pneumologia Ano de publicação: 2009 Tipo de documento: Artigo País de afiliação: Argentina Instituição/País de afiliação: Policlínica Bancaria/AR