Implante transcateter de valva aórtica sem a necessidade de marca-passo definitivo em uma série de casos consecutivos: é possível predizer o risco de bloqueio atrioventricular? / Transcatheter aortic valve implantatation without permanent pacemacker in a series of consecutive cases: is it possible to predict the risk of atrioventricular block?
Lemos, Pedro A; Mariane Júnior, José; Esteves Filho, Antonio; Kajita, Luiz J; Cardoso, Luiz F; Dracoulakis, Marianna D. A; Macedo, Ariane V. S; Mariño, Julio C. S; Jatene, Fabio B; Kalil Filho, Roberto.
Afiliação
  • Lemos, Pedro A; Hospital Sírio Libanês. São Paulo. BR
  • Mariane Júnior, José; Hospital Sírio Libanês. São Paulo. BR
  • Esteves Filho, Antonio; Hospital Sírio Libanês. São Paulo. BR
  • Kajita, Luiz J; Hospital Sírio Libanês. São Paulo. BR
  • Cardoso, Luiz F; Hospital Sírio Libanês. São Paulo. BR
  • Dracoulakis, Marianna D. A; Hospital Sírio Libanês. São Paulo. BR
  • Macedo, Ariane V. S; Hospital Sírio Libanês. São Paulo. BR
  • Mariño, Julio C. S; Hospital Sírio Libanês. São Paulo. BR
  • Jatene, Fabio B; Hospital Sírio Libanês. São Paulo. BR
  • Kalil Filho, Roberto; Hospital Sírio Libanês. São Paulo. BR
Rev. bras. cardiol. invasiva ; 18(2): 135-139, jun. 2010.
Article em Pt | LILACS | ID: lil-559918
Biblioteca responsável: BR44.1
RESUMO

Introdução:

Embora os stents farmacológicos tenham reduzido de forma inconteste o crescimento neointimal e sua consequência clínica direta, a reestenose, questionamentos acerca de sua segurança a longo prazo têm motivado o desenvolvimento de novas tecnologias. A aplicação sistêmica de fármacos antiproliferativos teria o potencial de solucionar o problema da reestenose e ao mesmo tempo amenizar a questão da segurança tardia inerente aos polímeros sintéticos. Recentemente, estudo investigando a administração intravenosa de uma solução de paclitaxel ligado à albumina (ABI-007) produziu efeito discreto na redução da formação de tecido neointimal. Este trabalho teve como objetivo avaliar, por meio da angiografia e do ultrassom intracoronário (USIC) seriados, a efetividade da aplicação intracoronária do ABI-007 na supressão da hiperplasia neointimal após implante de stents não-farmacológicos.

Método:

Entre novembro de 2006 e maio de 2007, pacientes portadores de lesões coronárias de novo, com extensão < 18 mm, em vasos nativos de 2,5 mm a 3,5 mm de diâmetro, foram submetidos a intervenção coronária percutânea com implante de stents não-farmacológicos, seguida de injeção intracoronária de 70 mg/m² de ABI-007. Avaliação com angiografia e USIC foi programada para todos os pacientes imediatamente após o procedimento e aos seis meses de seguimento. Desfecho primário era a perda luminal tardia (QCA) e o porcentual de obstrução volumétrica gerado pela hiperplasia neointimal (USIC) aos seis meses...
ABSTRACT

Background:

Transcatheter aortic valve implantation (TAVI) is increasingly reported as a valid alternative therapeutic option for patients with aortic valve stenosis with highsurgical risk. Although effective and relatively safe, TAVI is currently associated to the need of permanent pacemaker in 20% to 30% of the cases, with a number of predictors of this complication already described in the literature.

Method:

We report a series of 8 consecutive patients with aortic stenosis treated with TAVI (Corevalve Revalving, Medtronic Inc., USA).

Results:

The study population had ahigh surgical risk (on average, STS Score = 22.9% and EuroSCORE = 27.2%). All cases presented at least onehigh-risk predictor for post-procedure high grade atrioventricular block. The interventional procedure was successfully accomplished in all but one patient who presented a cardiac tamponade and in-hospital death. The transaortic pressure gradient was reduced from 91.3 ± 26.9 mmHg to19.4 ± 6.3 mmHg. After hospital discharge, the follow-up time ranged from 4 weeks to 12 months. During follow-up, there were no deaths and none of the patients had a new 2nd or 3rd degree atrioventricular block and permanent pacemaker was not required in any of the cases.

Conclusion:

Although the lack of a permanent pacemaker in the study population may have occurred by chance, our initial experience suggests that the need of pacemaker following TAVI is not readily predictable with the currently described high-risk predictors.
Assuntos
Palavras-chave
Texto completo: 1 Índice: LILACS Assunto principal: Estenose da Valva Aórtica / Marca-Passo Artificial / Próteses Valvulares Cardíacas Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: Pt Revista: Rev. bras. cardiol. invasiva Assunto da revista: CARDIOLOGIA / CIRURGIA GERAL Ano de publicação: 2010 Tipo de documento: Article
Texto completo: 1 Índice: LILACS Assunto principal: Estenose da Valva Aórtica / Marca-Passo Artificial / Próteses Valvulares Cardíacas Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: Pt Revista: Rev. bras. cardiol. invasiva Assunto da revista: CARDIOLOGIA / CIRURGIA GERAL Ano de publicação: 2010 Tipo de documento: Article