Comparação entre o manovacuômetro nacional e o importado para medida da pressão inspiratória nasal / Comparison between a national and a foreign manovacuometer for nasal inspiratory pressure measurement
Severino, Fernanda G; Resqueti, Vanessa R; Bruno, Selma S; Azevedo, Ingrid G; Vieira, Rudolfo H. G; Fregonezi, Guilherme A. F.
  • Severino, Fernanda G; s.af
  • Resqueti, Vanessa R; s.af
  • Bruno, Selma S; s.af
  • Azevedo, Ingrid G; s.af
  • Vieira, Rudolfo H. G; s.af
  • Fregonezi, Guilherme A. F; s.af
Braz. j. phys. ther. (Impr.) ; 14(5): 426-431, Sept.-Oct. 2010. graf, tab
Artigo em Português | LILACS | ID: lil-570723
RESUMO
CONTEXTUALIZAÇÃO A medida da pressão inspiratória nasal, conhecida como sniff teste, desenvolvida como um novo teste de força muscular inspiratória, utilizada principalmente em doenças neuromusculares, é de fácil realização e não invasiva. Apesar da importância clínica da avaliação da pressão inspiratória nasal, não existe um instrumento nacional disponível para realizá-la.

OBJETIVOS:

Comparar os manovacuômetros eletrônicos nacional e importado para a avaliação da pressão inspiratória nasal em pessoas saudáveis.

MÉTODOS:

Foram avaliados 18 voluntários saudáveis (idade 21,4±2,8 anos, IMC 23,4±2,5 Kg/m² , CVF 102,1±10,3 por centopred, VEF1 98,4±1 por centopred) por meio de duas medidas de pressão inspiratória nasal em dois equipamentos diferentes um nacional e outro importado. Todos os sujeitos realizaram a manobra no mesmo horário do dia, em dias ocasionais, sendo a ordem determinada aleatoriamente. Para análise estatística, foi utilizado o teste t pareado, a correlação de Pearson e o Bland-Altman com nível de significância de 5 por cento.

RESULTADOS:

As médias encontradas durante as duas medidas das pressões nasais foram de 125±42,4 cmH2O para o aparelho importado e de 131,7±28,7 cmH2O para o nacional. A análise de Pearson demonstrou uma correlação significativa entre as médias, com um coeficiente r=0,63. Os valores médios não apresentaram diferenças significativas pelo teste t pareado (p>0,05). Na análise de Bland-Altman, encontrou-se um BIAS igual a 7 cm H2O, desvio-padrão de 32,9 cmH2O para o DP e um intervalo de confiança de -57,5 cmH2O até 71,5 cmH2O.

CONCLUSÃO:

Os resultados encontrados sugerem que o manovacuômetro eletrônico nacional é viável e seguro para realização do sniff teste em sujeitos saudáveis.
ABSTRACT

BACKGROUND:

The measurement of nasal inspiratory pressure, known as the sniff test, was developed as a new test of inspiratory muscle strength, mainly used in neuromuscular conditions. The test is easy to be performed and noninvasive. Despite the clinical importance of assessment of nasal inspiratory pressure a national equipment is not available to assess it.

OBJECTIVES:

To compare a national with a foreign manovacuometer in assessing the nasal inspiratory pressure (sniff test) in healthy subjects.

METHODS:

18 subjects were evaluated (age 21.44±2.8 years, BMI 23.4±2.5 kg/m² , FVC 102.1±10.3 percent pred, FEV1 98.4±1 percent pred). We performed two measures of nasal inspiratory pressure using two different manovacuometer a national and a foreign. All subjects performed the tests at the same time of day, in different days being the order of the testes established randomly. It was used the paired t test, Pearson correlation and the Bland-Altman plots for statistical analysis considering a 5 percent significance level.

RESULTS:

The averages observed for the two measures of nasal pressures were 125±42.4 cmH2O for the foreign equipment, and 131.7±28.7 cmH2O for the national equipment. The Pearson correlation showed significant correlation between the means with a coefficient of r=0.63. The t test showed no significant differences between both measurements (p>0,05). The BIAS±SD found in Bland-Altman plot analysis was 7 cmH2O with limits of agreement between -57.5 cmH2O and 71.5 cmH2O.

CONCLUSION:

The results suggest that the national electronic device is feasible and safe to the sniff test measurement in healthy subjects.
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Testes de Função Respiratória / Músculos Respiratórios / Inalação / Força Muscular Tipo de estudo: Ensaio Clínico Controlado Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Braz. j. phys. ther. (Impr.) Assunto da revista: MEDICINA FISICA E REABILITACAO Ano de publicação: 2010 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Testes de Função Respiratória / Músculos Respiratórios / Inalação / Força Muscular Tipo de estudo: Ensaio Clínico Controlado Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Braz. j. phys. ther. (Impr.) Assunto da revista: MEDICINA FISICA E REABILITACAO Ano de publicação: 2010 Tipo de documento: Artigo