Correlación entre bilirrubina sérica y bilirrubinometría transcutánea en neonatos estratificados por edad gestacional / Correlation between serum bilirubin and transcutaneus bilirubinometry in newborns stratified by gestation age
Furzán, Jaime A; Expósito, Mercedes; Luchón, Consuelo.
  • Furzán, Jaime A; Hospital Universitario Doctor Alfredo Van Grieken. VE
  • Expósito, Mercedes; Hospital Universitario Doctor Alfredo Van Grieken. VE
  • Luchón, Consuelo; Hospital Universitario Doctor Alfredo Van Grieken. VE
Arch. venez. pueric. pediatr ; 70(2): 39-46, abr.-jun. 2007. tab, graf
Artigo em Espanhol | LILACS | ID: lil-589224
RESUMEN
La ictericia neonatal es un dilema diagnóstico y terapéutico muy común. Las decisiones basadas en mediciones seriadas de bilirrubina sérica se asocian con dolor en el niño, tiempo de espera y ansiedad por los resultados. La bilirrubinometría transcutánea puede ser una alternativa confiable si se demuestra su validez para poblaciones e instituciones particulares. Establecer el grado de correlación entre bilirrubina sérica total (BST) y bilirrubina transcutánea (BTC) en neonatos estratificados por edad gestacional. Estudio de diseño transversal. Cien recién nacidos ictéricos, 50 de término y 50 pretérmino, fueron analizados. A cada niño se le realizó una medición simultánea de BST y BTC. Luego se calcularon los coeficientes de variación de las medias, las diferencias de las medidas, la sensibilidad y el valor predictivo positivo de la BTC, los coeficientes de correlación, la ecuación de la línea de regresión, y la equivalencia de ambas mediciones para predecir el riesgo de hiperbilirrubinemia. El coeficiente de variación de las medias fue similar para ambas mediciones (BST 3,09 por ciento; BTC 3,24 por ciento). Las mediciones de BST y BTC fueron diferentes en 95 niños, e iguales en 5 de ellos. En los 95 neonatos con lecturas divergentes, la BTC subestimó la BST en 59 y la sobreestimó en 36, con una diferencia promedio de 1,39 mg/dl (DE 0,58 mg/dl). La sensibilidad de la BTC para identificar una BST >12 mg/dl fue 87 por ciento, con un valor predictivo positivo de 94 por ciento. El coeficiente de correlación de toda la muestra fue 0,88; en los neonatos de término fue 0,92 y en los pretérmino fue 0,84. El 82 por ciento de los neonatos fueron correctamente clasificados por la BTC en los percentiles de riesgo para hiperbilirrubinemia significativa. La precisión de las medidas de BST y BTC es similar en nuestra institución. Aunque hay una correlación aceptable entre ambos métodos de valoración, independientemente de la edad gestacional, la técnica de la BTC...
ABSTRACT
Newborn jaundice with its potential for producing brain damage remains a continuing problem for pediatricians. Therapeutic decisions based on serial measurements of serum bilirubin (TSB) are time-consuming and associated with stress for the child, parents and practitioners. Transcutaneous bilirubinometry (TCB) may be an alternative method if its reliability is proven for a particular institution and population. To establish the degree of correlation between TSB and TCB in newborn infants classified by gestational age. Cross-sectional study. 100 jaundiced infants, stratified into two groups of 50 preterm and 50 full-term newborns, were analyzed by simultaneous samples of TSB and TCB. Statistical analysis included estimates of differences between the two measurements, coefficients of variation of means, correlation coefficients, sensitivity and positive predictive value of TCB, and accuracy of TCB to predict significant hyperbilirubinemia. The coefficient of variation of means was similar for both methods (TSB 3.09 percent; TCB 3.24 percent). Measurements of TSB and TCB were different in 95 infants and equivalent in 5. TCB underestimated TSB in 59 measurements and overestimated it in 36 of them, with a mean difference of 1.39 mg/dl (SD 0.58 mg/dl). Sensitivity of TCB to identify a TSB >12 mg/dl was 87 percent, with a positive predictive value of 94 percent. The global correlation coefficient was 0.88 (term infants 0.92; preterm infants 0.84). 82 percent of the infants were correctly classified by the TCB within the percentiles of risk for significant hyperbilirubinemia. Precision of measurements of TSB and TCB is similar in our institution. Despite a good correlation between the two methods, regardless the gestational age, TCB tends to underestimate TSB when bilirubin values go beyond 12 mg/dl. The TCB technique can accurately predict the risk of extreme hyperbilirubinemia, so it can be cautiously used prior to dischange to reduce this risk and to avoid...
Assuntos

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Idade Gestacional / Hiperbilirrubinemia Neonatal / Kernicterus Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino / Recém-Nascido Idioma: Espanhol Revista: Arch. venez. pueric. pediatr Assunto da revista: Pediatria Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Venezuela Instituição/País de afiliação: Hospital Universitario Doctor Alfredo Van Grieken/VE

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Idade Gestacional / Hiperbilirrubinemia Neonatal / Kernicterus Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino / Recém-Nascido Idioma: Espanhol Revista: Arch. venez. pueric. pediatr Assunto da revista: Pediatria Ano de publicação: 2007 Tipo de documento: Artigo País de afiliação: Venezuela Instituição/País de afiliação: Hospital Universitario Doctor Alfredo Van Grieken/VE