Avaliação das condições de funcionamento dos equipamentos de raios X médico em serviços de radiologia no Estado da Paraíba, durante os anos de 2008 e 2009 / Evaluation of medical X-ray machines in Paraíba state radiology centers between 2008 and 2009
Radiol. bras
;
44(4): 244-248, jul.-ago. 2011. ilus, tab
Artigo
em Português
| LILACS
| ID: lil-598552
RESUMO
OBJETIVO:
Avaliar a coincidência entre o campo luminoso e o campo de radiação, e o alinhamento do feixe de radiação dos equipamentos de raios X médico no Estado da Paraíba, por meio de dois testes de controle de qualidade.MATERIAIS E MÉTODOS:
Foram utilizados os dispositivos para os testes de tamanho de campo e de alinhamento, um chassi carregado, trena e nível de bolha. Os testes de exatidão do sistema de colimação e de alinhamento do raio central do feixe de raios X foram realizados durante as inspeções sanitárias em serviços de radiologia nos anos de 2008 e 2009.RESULTADOS:
No ano de 2008, 121 equipamentos de raios X foram testados no Estado da Paraíba. No ano de 2009, passaram pelos testes 117 equipamentos. Deste universo, 86 foram selecionados para a comparação por terem sido avaliados tanto no ano de 2008 como em 2009, sendo observada uma melhoria de 18,60 por cento (n = 16) nos resultados dos testes realizados de um ano para outro. CONCLUSÃOConclusão:
Pode-se concluir que o percentual de equipamentos apresentando problemas no seu desempenho sofreu uma diminuição entre 2008 e 2009, não sendo observado programa de garantia de qualidade em nenhum dos serviços de radiologia do Estado da Paraíba.ABSTRACT
OBJECTIVE:
To evaluate the light and radiation fields congruence and radiation beam alignment in medical X-ray equipment in Paraíba state radiology centers by means of two quality control tests. MATERIALS ANDMETHODS:
A loaded cassette, a measuring tape and a bubble level were utilized in the field size and alignment testing. The evaluation of collimation systems accuracy and X-ray beam alignment was undertaken during health inspections performed in radiology centers between 2008 and 2009.RESULTS:
In 2008, 121 X-ray machines were evaluated in the Paraíba state. In 2009, 117 machines were tested. From this universe, 86 machines were selected for comparison, since they were evaluated both in 2008 and 2009, with 18.60 percent (n = 16) showing test results improvement from one year to another.CONCLUSION:
The percentage of problematic X-ray machines decreased between 2008 and 2009, notwithstanding no quality assurance program has been observed in Paraíba state radiology centers.
Texto completo:
DisponíveL
Índice:
LILACS (Américas)
Assunto principal:
Controle de Qualidade
/
Diagnóstico por Imagem
/
Radiografia
/
Equipamentos para Diagnóstico
Tipo de estudo:
Estudo diagnóstico
/
Estudos de avaliação
/
Estudo observacional
País/Região como assunto:
América do Sul
/
Brasil
Idioma:
Português
Revista:
Radiol. bras
Assunto da revista:
Diagnóstico por Imagem
/
Radiologia
Ano de publicação:
2011
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Agência Estadual de Vigilância Sanitária da Paraíba/BR