Síndrome de Twiddler: una causa infrecuente de disfunción de marcapaso definitivo. Caso clínico / Twiddler syndrome: Report of one case
Rev. méd. Chile
;
139(7): 917-919, jul. 2011. ilus
Artigo
em Espanhol
| LILACS
| ID: lil-603146
ABSTRACT
We report a 26year old patient who had a single chamber pacemaker implantation one year before. During a routine pre-operative evaluation, pacemaker dysfunction was demonstrated due to sensing and pacing failure, associated to left pectoral muscle rhythmic contraction. Chest X-ray confirmed Twiddler syndrome, in which twisting or rotation of the device inside the pocket results in lead dislodgement and device malfunction.