Utilidad del Pentacam para medir el poder corneal después de cirugía refractiva con excímer láser / Use of Pentacam camera in measuring the corneal power after Excimer laser refractive surgery
Cuan Aguilar, Yoriel; Pérez Candelaria, Eneida de la C; Montero Díaz, Eric; Santiesteban García, Imalvet; Ortega Díaz, Lorelei; Cárdenas Díaz, Taimí.
  • Cuan Aguilar, Yoriel; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Pérez Candelaria, Eneida de la C; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Montero Díaz, Eric; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Santiesteban García, Imalvet; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Ortega Díaz, Lorelei; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Cárdenas Díaz, Taimí; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
Rev. cuba. oftalmol ; 23(supl.1): 513-521, 2010.
Artigo em Espanhol | LILACS | ID: lil-615588
RESUMEN

OBJETIVO:

Evaluar el poder corneal brindado por el Pentacam en pacientes miopes operados de cirugía refractiva con excimer láser en el Instituto Cubano de Oftalmología Ramón Pando Ferrer, desde septiembre a diciembre de 2008.

MÉTODOS:

Se realizó un estudio prospectivo en 50 ojos de 25 pacientes, utilizando varias mediciones obtenidas por el Pentacam mapas de color queratométrico total y Equivalent K-reading Power, el true net power y las lecturas de queratometría equivalentes a diferentes diámetros. Todas analizadas y comparadas con los valores de queratometrías obtenidos a través del método de historia clínica. Además se compararon las queratometrías preoperatorias reales con las queratometrías preoperatorias estimadas mostradas por el Pentacam. El análisis estadístico se realizó con la Prueba T para datos pareados, utilizando una significación del 95 por ciento.

RESULTADOS:

No hubo diferencias estadísticas significativas entre las queratometrías del método de historia clínica, el Equivalent K-reading Power y las lecturas de queratometría equivalentes a diferentes diámetros, mostrando la de 3,0 mm, la menor diferencia. El resto de las mediciones difirieron de forma significativa. No se encontró diferencias entre las queratometrías preoperatorios.

CONCLUSIONES:

El Pentacam aportó poderes corneales que no difirieron estadísticamente de los obtenidos por el método de historia clínica en ojos miopes que tenían cirugía previa con excímer láser
ABSTRACT

OBJECTIVE:

To evaluate the corneal power offered by Pentacam for myopic patients operated on by refractive excimer laser surgery at Ramón Pando Ferrer Cuban Institute of Ophthalmology from September to December, 2008.

METHODS:

A prospective study was performed in 50 eyes of 25 patients, using several measurements taken by Pentacam total keratometric color maps and Equivalent K-reading Power, true net power and the keratometry readings equivalent to different diameters. All these measurements were analyzed and compared with the keratometry values seen in the clinical records. The real preoperative keratometries were also compared with the estimated preoperative keratometries shown by the Pentacam. The statistical analysis was based on paired t test, with 95 percent significance.

RESULTS:

There were not significant statistical differences among the Keratometries of the medical histories, the Equivalent K-reading Power and the Keratometry readings equivalent to different diameters, being the 3.0 mm reading the one that showed the smallest difference. The rest of the measurements significantly differed. There were no differences among the preoperative keratometries.

CONCLUSIONS:

The Pentacam camera provided corneal powers that did not differ statistically from those found in the medical history in the myopic eyes that had previously undergone Excimer laser surgery
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Córnea / Topografia da Córnea / Lasers de Excimer Tipo de estudo: Estudo observacional Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2010 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología Ramón Pando Ferrer/CU

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Córnea / Topografia da Córnea / Lasers de Excimer Tipo de estudo: Estudo observacional Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2010 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología Ramón Pando Ferrer/CU