Comparação do tempo de dois procedimentos com novas tecnologias de Potencial Evocado Auditivo de Tronco Encefálico Automático (PEATE-A) / Testing time comparison between two procedures with new technologies of Automated Auditory Brainstem Response (AABR)
CoDAS
; 25(1): 34-38, 2013. ilus, tab
Article
em Pt
| LILACS
| ID: lil-672155
Biblioteca responsável:
BR1.1
RESUMO
OBJETIVO:
Comparar o tempo de dois procedimentos de Potencial Evocado Auditivo de Tronco Encefálico Automático (PEATE-A) utilizando taxa de repetição e modo de detecção diferentes.MÉTODOS:
Foi realizada a triagem auditiva com Potencial Evocado Auditivo de Tronco Encefálico Automático (PEATE-A) em 30 neonatos, com média de idade de 21 dias. Cada neonato foi submetido a dois procedimentos de triagem auditiva com diferentes equipamentos de PEATE-A (Procedimento 1 e Procedimento 2). O Procedimento 1 utilizou taxa de repetição de 53 Hz e teste estatístico para detecção da resposta denominado one-sample test, e o Procedimento 2 utilizou taxa de repetição de 90 Hz e teste estatístico para detecção da resposta denominado q-sample test. Em seguida, foi registrado o PEATE com estímulo clique como teste padrão ouro, cuja análise das respostas foi realizada por um fonoaudiólogo com experiência neste procedimento.RESULTADOS:
O tempo médio observado para o Procedimento 1 considerando ambas orelhas foi de 84,8 (±53,5) segundos; para o Procedimento 2, o tempo médio foi de 27,9 (±20,0) segundos. O primeiro procedimento apresentou tempo de exame aproximadamente três vezes maior quando comparado ao segundo. A análise estatística mostrou diferença significativa entre os tempos dos procedimentos.CONCLUSÃO:
O q-sample test e taxa de repetição em torno de 90 Hz nos equipamentos de PEATE-A detectaram a resposta de maneira mais rápida.ABSTRACT
PURPOSE:
To compare the testing time of two Automated Auditory Brainstem Response (AABR) procedures using different repetition rates and detection modes.METHODS:
A hearing screening using AABR was performed in 30 newborns with mean age of 21 days. Each newborn was submitted to two hearing screening procedures using different AABR equipments (Procedure 1 and Procedure 2). Procedure 1 used a repetition rate of 53 Hz and the one-sample test for response detection; Procedure 2 used a repetition rate of 90 Hz, and the q-sample test for response detection. The ABR with click stimulus was then registered as gold standard test, and the responses were analyzed by a trained audiologist.RESULTS:
The mean time observed for Procedure 1 considering both ears was 84.8 (±53.5) seconds; for Procedure 2 the mean time was 27.9 (±20.0) seconds. The testing time of the first procedure was three times longer than the second one. Statistical analysis showed significant difference between the testing times of the procedures.CONCLUSION:
The q-sample test and the repetition rate of 90 Hz used in the AABR equipment showed earlier response detection.Palavras-chave
Texto completo:
1
Índice:
LILACS
Assunto principal:
Potenciais Evocados Auditivos do Tronco Encefálico
/
Triagem Neonatal
/
Transtornos da Audição
/
Testes Auditivos
Tipo de estudo:
Observational_studies
/
Prevalence_studies
/
Risk_factors_studies
Limite:
Humans
/
Newborn
Idioma:
Pt
Revista:
CoDAS
Assunto da revista:
AUDIOLOGIA
/
PATOLOGIA DA FALA E LINGUAGEM
Ano de publicação:
2013
Tipo de documento:
Article