Screw loosening of different UCLA-type abutments after mechanical cycling
Junqueira, Marcela C; Ribeiro, Ricardo F; Faria, Adriana Claudia L; Macedo, Ana Paula; Almeida, Rossana P.
  • Junqueira, Marcela C; University of São Paulo. Dental School of Ribeirão Preto. Department of Dental Materials and Prosthodontics. Ribeirão Preto. BR
  • Ribeiro, Ricardo F; University of São Paulo. Dental School of Ribeirão Preto. Department of Dental Materials and Prosthodontics. Ribeirão Preto. BR
  • Faria, Adriana Claudia L; University of São Paulo. Dental School of Ribeirão Preto. Department of Dental Materials and Prosthodontics. Ribeirão Preto. BR
  • Macedo, Ana Paula; University of São Paulo. Dental School of Ribeirão Preto. Department of Dental Materials and Prosthodontics. Ribeirão Preto. BR
  • Almeida, Rossana P; University of São Paulo. Dental School of Ribeirão Preto. Department of Dental Materials and Prosthodontics. Ribeirão Preto. BR
Braz. j. oral sci ; 12(3): 228-232, July-Sept. 2013. ilus, tab
Artigo em Inglês | LILACS | ID: lil-701311
ABSTRACT

AIM:

To evaluate the loss of applied torque (detorque) values in cast and pre-machined abutments for external hex abutment/implant interface of single implant-supported prostheses subjected to mechanical cycling.

METHODS:

Ten metal crowns were fabricated using two types of UCLA abutments cast and pre-machined with metal base in NiCrTi alloy and tightened to regular external hex implants with a titanium alloy screw, with an insertion torque of 32 N.cm, measured with a digital torque gauge. Samples were embedded with autopolymerizing acrylic resin in a stainless steel cylindrical matrix, and positioned in an electromechanical machine. Dynamic oblique loading of 120 N was applied during 5 x 10(5) cycles. Then, each sample was removed from the resin and detorque values were measured using the same digital torque gauge. The difference of the initial (torque) and final (detorque) measurement was registered and the results were expressed as percentage of initial torque. The results of torque loss were expressed as percentage of the initial torque and subjected to statistical analysis by the Student's t-test (p<0.05) for comparisons between the test groups.

RESULTS:

Statistical analysis demonstrated that mechanical cycling reduced the torque of abutments without significant difference between cast or pre-machined UCLA abutments (p=0.908).

CONCLUSIONS:

Within the limitations of this in vitro study, it may be concluded that the mechanical cycling, corresponding to one-year use, reduced the torque of the samples regardless if cast or pre-machined UCLA abutments were used.
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Titânio / Dente Suporte / Torque / Implantação Dentária Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2013 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: University of São Paulo/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Titânio / Dente Suporte / Torque / Implantação Dentária Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2013 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: University of São Paulo/BR