Configuración del Pentacam en el tamizaje de cirugía refractiva / Pentacam configuration in refractive surgery screening
Sánchez Bravo, Justo; Riverón Ruíz, Yanaisa; Ortega Díaz, Lorelei; Rodríguez Sánchez, Sorania; Miranda Hernández, Iramis; Lantigua Maldonado, Isabel.
  • Sánchez Bravo, Justo; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Riverón Ruíz, Yanaisa; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Ortega Díaz, Lorelei; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Rodríguez Sánchez, Sorania; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Miranda Hernández, Iramis; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
  • Lantigua Maldonado, Isabel; Instituto Cubano de Oftalmología Ramón Pando Ferrer. La Habana. CU
Rev. cuba. oftalmol ; 26(3): 417-426, sep.-dic. 2013.
Artigo em Espanhol | LILACS | ID: lil-706673
RESUMEN

Objetivo:

determinar la consistencia de la configuración manual del Pentacam en el tamizaje de cirugía refractiva.

Métodos:

se realizó un estudio observacional, descriptivo en el Instituto Cubano de Oftalmología Ramón Pando Ferrer durante el 2012. La muestra quedó conformada por 104 imágenes correspondiente a ojos examinados por el Pentacam que cumplieron los criterios de selección. Se analizaron variables tales como diámetro para al cálculo de la esfera de mejor ajuste, elevación anterior, elevación posterior y se utilizó la clasificación de ectasias corneales según elevación anterior y posterior para agrupar la muestra.

Resultados:

del total de la muestra el 45,2 por ciento y el 54,8 por ciento utilizó un diámetro menor de 8 mm, y mayor de 8 mm para el cálculo de la esfera de mejor ajuste, respectivamente. La comparación de media de los dos grupos fue estadísticamente significativa. Según la clasificación de ectasia por elevación anterior y posterior, con un diámetro mayor de 8 mm se duplican en por ciento los exámenes normales y disminuyen los patológicos en un 25 porciento respectivamente. Existen diferencias entre las medias y la desviación estándar de los exámenes evolutivos a un mismo ojo según la configuración automática y manual.

Conclusiones:

la configuración manual del Pentacam permite tener una consistencia en el tamizaje de cirugía refractiva
ABSTRACT

Objective:

to determine the consistency of manual Pentacam configuration in refractive surgery screening.

Methods:

observational and descriptive study performed at Ramon Pando Ferrer Cuban Institute of Ophthalmology during 2012. The sample was made up of 104 images of eyes examined by Pentacam equipment, which had met the selection criteria. The analyzed variables were diameter for best fit sphere calculation, anterior elevation, posterior elevation and the classification of corneal ectasias based on anterior and posterior elevations was used in order to group the sample.

Results:

in the sample, 45.2 percent and 54.8 percent used a diameter lower than 8mm and over 8mm for the calculation of the best fit sphere, respectively. The mean comparison between the two groups was statistically significant. According to the ectasia classification based on anterior and posterior elevations, with the over 8mm diameter, the normal exams doubled in some percentage and the pathological ones decreased by 25 percent, respectively. There are differences between the means and the standard deviations in evolutional exams performed in the same eye, considering the automatic and the manual configurations.

Conclusions:

manual configuration of Pentacam allows attaining consistency in the refractive surgery screening
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Peneiramento de Líquidos / Procedimentos Cirúrgicos Refrativos Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Fatores de risco / Estudo de rastreamento Limite: Humanos Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2013 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología Ramón Pando Ferrer/CU

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Peneiramento de Líquidos / Procedimentos Cirúrgicos Refrativos Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Fatores de risco / Estudo de rastreamento Limite: Humanos Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2013 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología Ramón Pando Ferrer/CU