Your browser doesn't support javascript.
loading
Avaliação da retenção mecânica em implantes cilíndricos e cônicos quando submetidos ao teste de tração: estudo in vitro / Evaluation of mechanical retention in cylindrical and tapered implants when subjected to tensile strength test: in vitro study
Silva, Fernando dos Santos Pereira; Lima, Waldir Benincasa de Castro; Paço, Franceni Durço.
  • Silva, Fernando dos Santos Pereira; UniFOA. BR
  • Lima, Waldir Benincasa de Castro; ABOVR. BR
  • Paço, Franceni Durço; ABOVR. BR
ImplantNews ; 11(1): 79-84, 2014. ilus, tab
Artigo em Português | LILACS, BBO | ID: lil-730043
RESUMO
A estabilidade primária dos implantes dentários influencia a permanência desses em função, exercendo papel preponderante no índice de sucesso. Sabe-se que a sobrevida do implante não depende unicamente da estabilidade primária, mas nos procedimentos de carga imediata, principalmente, ela é fundamental.

Objetivos:

no presente trabalho, testou-se a capacidade de retenção mecânica de dois tipos de implantes com a forma cilíndrica (3,75 x 11,0 mm) e com a forma cônica (4,0 x 11,0 mm), inseridos em resina de poliuretano e conforme as instruções dos fabricantes. Material e

métodos:

os torques de inserção foram determinados com um torquímetro e a remoção dos implantes foi realizada com o auxílio de dinamômetro especialmente montado para o experimento.

Resultados:

as médias de resistência à tração para os implantes cônicos e cilíndricos foram 497 N e 366 N, respectivamente. A análise estatística acusou diferença significante entre os grupos testados (p < 0,05).

Conclusão:

1) implantes cônicos tiveram melhor desempenho, quanto ao quesito retenção mecânica, que os implantes cilíndricos, 2) implantes cônicos necessitaram do maior torque aferido no ato de sua inserção quando comparados aos implantes cilíndricos.
ABSTRACT
Primary dental implant stability influences its function being decisive for success rates. It is well-known that implant survival is not only influenced by this issue, but under immediate loading procedure this aspect is mandatory.

Objectives:

in this paper, the mechanical retention of two dental implant geometries cylindrical (3.75 x 10 mm) and a tapered (4.0 x 10 mm) was tested with both systems inserted into polyurethane resin bodies according to manufacturer´s recommendations. Material and

methods:

the final insertion values were registered with a torquemeter and all implants were removed under tensile strength with the aid of a dynamometer.

Results:

mean values for cylindrical and tapered geometries were 497 N and 366 N, respectively. The statistical analysis demonstrated significant differences at 5% level.

Conclusion:

1) tapered implants demonstrated better performance regarding tensile test than cylindrical implants, 2) tapered implants demonstrated higher torque levels when compared to cylindrical implants.
Assuntos

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Próteses e Implantes / Teste de Materiais Tipo de estudo: Guia de Prática Clínica Idioma: Português Revista: ImplantNews Assunto da revista: Odontologia Ano de publicação: 2014 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: ABOVR/BR / UniFOA/BR

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Próteses e Implantes / Teste de Materiais Tipo de estudo: Guia de Prática Clínica Idioma: Português Revista: ImplantNews Assunto da revista: Odontologia Ano de publicação: 2014 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: ABOVR/BR / UniFOA/BR