Chemical analysis and Vickers hardness of orthodontic mini-implants
Feitoza, Christiane Cavalcante; Oliveira, Guilherme José Pimentel Lopes de; Consani, Rafael Leonardo Xediek; Boeck, Eloisa Marcantonio; Pizzol, Karina Eiras Dela Coleta; Lunardi, Nadia.
  • Feitoza, Christiane Cavalcante; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
  • Oliveira, Guilherme José Pimentel Lopes de; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
  • Consani, Rafael Leonardo Xediek; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
  • Boeck, Eloisa Marcantonio; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
  • Pizzol, Karina Eiras Dela Coleta; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
  • Lunardi, Nadia; Centro Universitário de Araraquara. Department of Dentistry. Araraquara. BR
Braz. j. oral sci ; 14(1): 1-4, Jan-Mar/2015. tab, graf
Artigo em Inglês | LILACS | ID: lil-745788
ABSTRACT
Orthodontic mini-implants are used in clinical practice to provide efficient and aesthetically-pleasing anchorage.

AIM:

To evaluate the hardness Vickers hardness and chemical composition of mini-implant titanium alloys from five commercial brands.

METHODS:

Thirty self-drilling mini-implants, six each from the following commercial brands, were used Neodent NEO, Morelli MOR, Sin SIN, Conexão CON, and Rocky Mountain RMO. The hardness and chemical composition of the titanium alloys were performed by the Vickers hardness test and energy dispersive X-ray spectroscopy, respectively.

RESULTS:

Vickers hardness was significantly higher in SIN implants than in NEO, MOR, and CON implants. Similarly, VH was significantly higher in RMO implants than in MOR and NEO ones. In addition, VH was higher in CON implants than in NEO ones. There were no significant differences in the proportions of titanium and aluminum in the mini-implant alloy of the five commercial brands. Conversely, the proportion of vanadium differed significantly between CON and MOR/NEO implants.

CONCLUSIONS:

Mini-implants of different brands presented distinct properties of hardness and composition of the alloy...
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Ortodontia / Titânio / Materiais Dentários / Fenômenos Químicos / Dureza Limite: Humanos Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2015 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro Universitário de Araraquara/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Ortodontia / Titânio / Materiais Dentários / Fenômenos Químicos / Dureza Limite: Humanos Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2015 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Centro Universitário de Araraquara/BR