Your browser doesn't support javascript.
loading
Mechanical analysis of femoral neck fracture fixation with dynamic condylar screw in synthetic bone
Freitas, Anderson Freitas; Maciel, Rafael Almeida; Lima, Renato de Almeida; Souto, Diogo Ranier de Macedo; Ferrer, Marcelo de Almeida.
  • Freitas, Anderson Freitas; Hospital Ortopédico e Medicina Especializada. Brasília. BR
  • Maciel, Rafael Almeida; Hospital Ortopédico e Medicina Especializada. Brasília. BR
  • Lima, Renato de Almeida; Hospital Ortopédico e Medicina Especializada. Brasília. BR
  • Souto, Diogo Ranier de Macedo; Hospital Ortopédico e Medicina Especializada. Brasília. BR
  • Ferrer, Marcelo de Almeida; Hospital Ortopédico e Medicina Especializada. Brasília. BR
Acta ortop. bras ; 22(5): 264-268, Sep-Oct/2014. ilus, tab, graf
Artigo em Inglês, Português | LILACS | ID: lil-783280
RESUMO
Analisar estatisticamente resultados obtidos em ensaiosbiomecânicos da fixação de fratura do colo do fêmur Pauwelstipo III, em osso sintético, com parafuso dinâmico condilar (DCS) egrupo controle.

Método:

Foram utilizados dez ossos sintéticos defabricação nacional. Grupo teste foi realizada fixação após osteotomiaa 70o de inclinação, utilizando placa DCS de quatro furos.Analisou-se a resistência desta fixação em 5 mm de deslocamentoe desvio rotacional (Fase 1), e a 10 mm (Fase 2). Grupo controle osmodelos foram ensaiados em sua integridade até que ocorresse afratura do colo femoral.

Resultados:

Os valores do ensaio no grupoteste na fase 1 apresentaram média de 974N e DP= 114N. Na fase2, obtiveram média de 1335N e DP=98N. Os valores da carga nogrupo controle foram 1544N, 1110N, 1359N, 1194N, 1437N, respectivamente.A análise estatística pelo teste de Mann-Whitney paracomparação da força máxima (N) entre o grupo teste e o controle,na fase 2, demostrou que não existe diferença significativa entre asplacas DCS e controle (p = 0,91).

Conclusão:

Não existe diferençasignificativa entre as placas DCS e o grupo controle submetido aforça máxima. Nível de Evidência III, Caso Controle...
ABSTRACT
To analyze statistically results in biomechanical testingof fixation of femoral neck Pauwels type III fractures, on syntheticbone, with dynamic condylar screw (DCS) and control group.

Methods:

Ten synthetic bones of a national brand were used. TestGroup fixation was performed after osteotomy at 70o tilt usingDCS plate with four holes. We analyzed the resistance of thisfixation with 5 mm displacement and rotational deviation (Step 1)and with10 mm (Step 2). Control group the models were testedin their integrity until the femoral neck fracture occurred.

Results:

The values of the test group in Step 1 showed a mean of 974Nand SD = 114N. In Stage 2, we obtained on average 1335N andSD = 98N. The values in the control group were 1544N, 1110N,1359N, 1194N, 1437N, respectively. Statistical analysis using theMann-Whitney test for comparison of the maximum force (N) betweenthe test group and the control, in Step 2, demonstrated thatthere is no significant difference between the DCS and controlplates (p = 0.91).

Conclusion:

There is no significant differencebetween the DCS boards and the control group exposed to fullresistance. Level of Evidence III, Case Control...
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Osteotomia / Fenômenos Biomecânicos / Osso e Ossos / Parafusos Ósseos / Fixadores Internos / Fraturas do Colo Femoral / Fixação de Fratura Idioma: Inglês / Português Revista: Acta ortop. bras Assunto da revista: Ortopedia Ano de publicação: 2014 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Hospital Ortopédico e Medicina Especializada/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Osteotomia / Fenômenos Biomecânicos / Osso e Ossos / Parafusos Ósseos / Fixadores Internos / Fraturas do Colo Femoral / Fixação de Fratura Idioma: Inglês / Português Revista: Acta ortop. bras Assunto da revista: Ortopedia Ano de publicação: 2014 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Hospital Ortopédico e Medicina Especializada/BR