Your browser doesn't support javascript.
loading
Comparison of simultaneous estimation of cardiac output by four techniques in patients undergoing off-pump coronary artery bypass surgery--a prospective observational study.
Ann Card Anaesth ; 2007 Jul; 10(2): 121-6
Artigo em Inglês | IMSEAR | ID: sea-1404
ABSTRACT
We prospectively compared four techniques of cardiac output measurement bolus thermodilution cardiac output (TDCO), continuous cardiac output (CCO), pulse contour cardiac output (PiCCO), and Flowtrac (FCCO), simultaneously in fifteen patients undergoing off-pump coronary artery bypass grafting (OPCAB). All the patients received pulmonary artery catheter (capable of measuring both bolus thermodilution cardiac output and CCO), PiCCO arterial cannula in the left and FCCO in the right femoral artery. Cardiac indices (CI) were obtained every fifteen minutes by using all the four techniques. TDCO was treated as 'control' and the rest were treated as 'test' values. Interchangeability of techniques with TDCO was assessed by Bland and Altman plotting and mountain plot. Four hundred and thirty eight sets of data were obtained from fifteen patients. The values of cardiac output varied between 1 to 6.9 L/min. We found that the values of all the techniques were interchangeable. At certain times, the values of CI measured by both PiCCO and FCCO appeared erratic. The values of CI measured simultaneously appeared in the following descending order of accuracy; TDCO>CCO>FCCO>PiCCO (the % times TDCO correlated with CCO, FCCO, PiCCO was 93, 86 and 80 respectively). The bias and precision (in L/min) for CCO were 0.03, 0.06, PiCCO 0.13, 0.1 and flowtrac 0.15, 0.04 respectively suggesting interchangeability. We conclude that the cardiac output measured by CCO technique and the pulse contour as measured by PiCCO and FCCO were interchangeable with TDCO more than 80% of the times.
Assuntos
Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Pulso Arterial / Termodiluição / Feminino / Humanos / Masculino / Cateterismo de Swan-Ganz / Débito Cardíaco / Estudos Prospectivos / Reprodutibilidade dos Testes / Monitorização Intraoperatória Tipo de estudo: Estudo observacional Idioma: Inglês Revista: Ann Card Anaesth Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Pulso Arterial / Termodiluição / Feminino / Humanos / Masculino / Cateterismo de Swan-Ganz / Débito Cardíaco / Estudos Prospectivos / Reprodutibilidade dos Testes / Monitorização Intraoperatória Tipo de estudo: Estudo observacional Idioma: Inglês Revista: Ann Card Anaesth Ano de publicação: 2007 Tipo de documento: Artigo