Pressure bandage over venous conduit harvesting site causing compartment syndrome in a patient with intra-aortic balloon pump: An unusual cause.
Ann Card Anaesth ; 2015 Jul; 18(3): 453-459
Artigo em Inglês | IMSEAR | ID: sea-162402
ABSTRACT
Intra Aortic Balloon Pump (IABP) is conventionally used to support coronary perfusion and weaning from cardiopulmonary bypass. IABP in situ has its own share of complications. We present a case where a patient on IABP support who had reduced peripheral pulsations of the ipsilateral limb and was initially misdiagnosed as IABP catheter associated thromboembolism. A negative embolectomy ruled out the same. Further looking for the cause of reduction of ipsilateral pulses it was found that the tight compressive bandage at saphenous vein conduit harvesting site had led to development of compartment syndrome (CS).
Assuntos

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Pressão / Veia Safena / Tromboembolia / Humanos / Masculino / Prótese Vascular / Síndromes Compartimentais / Adulto / Coleta de Tecidos e Órgãos / Bandagens Compressivas Idioma: Inglês Revista: Ann Card Anaesth Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Pressão / Veia Safena / Tromboembolia / Humanos / Masculino / Prótese Vascular / Síndromes Compartimentais / Adulto / Coleta de Tecidos e Órgãos / Bandagens Compressivas Idioma: Inglês Revista: Ann Card Anaesth Ano de publicação: 2015 Tipo de documento: Artigo