Your browser doesn't support javascript.
loading
Comparison of I-gel and ProSeal laryngeal mask airway during volume controlled ventilation
Artigo | IMSEAR | ID: sea-187071
ABSTRACT

Introduction:

I-gel and the ProSeal laryngeal mask airway (PLMA) are two supra-glottic airway devices with gastric channel used for airway maintenance in anesthesia. This study was designed to evaluate the efficacy of I-gel compared with PLMA for airway maintenance in patients under general anesthesia with controlled ventilation. The aim of the study To compare the supra-glottic airway devices, I-Gel and ProSeal Laryngeal mask airway with respect to Ease of insertion, Time taken for insertion, Airway leak pressure, Hemodynamic response during intubation. Materials and

methods:

A total number of, 40 patients were randomized into two groups of 20 each. After induction of anesthesia using a standardized protocol for all the patients, one of the supra-glottic airway devices was inserted. Insertion parameters, ease of gastric tube insertion, airway leak pressure, hemodynamic changes, were noted.

Results:

There was no significant difference in the incidence of adverse effects in both the groups. One incidence of airway trauma was noted in I-gel group. No gastric insufflations and laryngo or bronchospasm in both groups.

Conclusion:

Based on the result of our study we conclude that I-gel had an acceptable airway leak pressure of 23 cm H2O when compared to ProSeal whose airway leak pressure is significantly higher i-e 29 cm H2O.Both the devices provided optimal oxygenation and no fall in saturation was observed in both the groups.

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Tipo de estudo: Ensaio Clínico Controlado Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Tipo de estudo: Ensaio Clínico Controlado Ano de publicação: 2018 Tipo de documento: Artigo