Use of Ionic Liquids for Preconcentration/Separation of Diazinon by In Situ Solvent Formation Microextraction (ISFME) and Measurement by the UV-Vis Spectrophotometer
Artigo | IMSEAR | ID: sea-203663
ABSTRACT
In the current study, the UV-Vis spectrophotometer has been used for measuring the Diazinon. The extraction andpreconcentration of Diazinon was done by the use of In Situ Solvent Formation Microextraction (ISFME) method, and theionic liquids 1-Hexyl 3-methyl-imidazolium bromide [Hmim] [Br] and lithium (bis-trifluoromethylsulfonyl) -imide [Li(NTF2)] as the opposite ion. Different parameters effective on extraction such as the sample solvent pH, the amount of ionicliquid, the centrifuge duration, the amount of salt, and the amount of the opposite ion (NTF2-) were investigated at the twoscreening and response levels by the experiment designing software Statgraphic X64, in order to obtain the optimalconditions. Based on the obtained results, the highest rate of extraction was obtained as 95.41% in the pH of -11.77 for 1 mlsolvent of 0.5 mg/Lit containing Diazinon, 120 mg of ionic liquid, 1 ml of the opposite ion with a concentration of 0.02molar, and the duration of 5 minutes in the centrifuge. The limit of detection of this method is 0.42 mg/Lit based on the 3Sbequation and 0.69 mg/Lit based on the 3Sb/m equation. The relative standard deviation is 1.76% and the enrichment factorwas calculated as 48.66.

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Ano de publicação: 2019 Tipo de documento: Artigo