Your browser doesn't support javascript.
loading
Long term performance of silicone insulated and polyurethane insulated cardiac pacing leads.
Indian Heart J ; 1992 May-Jun; 44(3): 145-9
Artigo em Inglês | IMSEAR | ID: sea-3058
ABSTRACT
The long term performance of various pacing leads in use for at least one year is reported from one centre. Between January 1975 and December 1990, 1056 cardiac pacing leads were implanted in 881 patients (mean age 57 +/- 13 years). Eight hundred and three leads were silicone insulated (SI) and 253 leads were polyurethane insulated (PUI). Leads from different manufacturers were used (Medtronic, CPI, Telectronics). The average duration of follow up was 56.6 +/- 77 months (range 12-221 months) for SI and 47.1 +/- 24 months (range 12-99 months) for PUI leads. Overall lead failure occurred more often in the PUI group (n = 46, 18.2%) compared to SI group (n = 93, 11.3%, p < 0.025). Majority of lead failures occurred in the first 36 months after implantation. In particular, Medtronic models 6971, 6972 (urethane) and 6901 (silicone) showed higher rate of complications (25.5%, 26.3% and 44.5% respectively).
Assuntos
Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Marca-Passo Artificial / Poliuretanos / Síndrome do Nó Sinusal / Silicones / Idoso de 80 Anos ou mais / Idoso / Feminino / Humanos / Masculino / Criança Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Aged80 Idioma: Inglês Revista: Indian heart j Ano de publicação: 1992 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Marca-Passo Artificial / Poliuretanos / Síndrome do Nó Sinusal / Silicones / Idoso de 80 Anos ou mais / Idoso / Feminino / Humanos / Masculino / Criança Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Aged80 Idioma: Inglês Revista: Indian heart j Ano de publicação: 1992 Tipo de documento: Artigo