Limitation of glucose oxidase method of glucose estimation in jaundiced neonates.
Indian J Exp Biol ; 2006 Mar; 44(3): 254-5
Artigo em Inglês | IMSEAR | ID: sea-62658
ABSTRACT
The most widely used method for estimation of plasma glucose is that adopted by Trinder's using glucose oxidase-peroxidase (GOD-POD) system. This method gives much lower blood glucose values with blood samples of neonatal jaundice (plasma bilirubin level > 10 mg/dL) of age 10 +/- 5 daysthan with samples of neonates of the same age group without jaundice or older children suffering from other diseases like acute respiratory distress, septicemia.