Validation of ESR analyzer using Westergren ESR method.
Indian J Pathol Microbiol ; 2007 Jul; 50(3): 634-5
Artigo em Inglês | IMSEAR | ID: sea-72727
ABSTRACT
Erythrocyte sedimentation rate (ESR) is one of the most frequently ordered laboratory test. ESR analyzers were developed to provide a quick and efficient measure of ESR. We compared the results of ESR obtained by an ESR analyzer with those by the Westergren method in a group of 75 patients Linear regression analysis showed a good correlation between the two results (r = 0.818, p < 0.01). The intra class correlation was 0.82. The analyzer method had the advantages of safety, decreased technician time and improved patient care by providing quick results.
Assuntos
Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Autoanálise / Vácuo / Sedimentação Sanguínea / Idoso / Feminino / Humanos / Masculino / Modelos Lineares / Criança / Pré-Escolar Tipo de estudo: Estudos de avaliação / Estudo prognóstico Idioma: Inglês Revista: Indian J Pathol Microbiol Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: IMSEAR (Sudeste Asiático) Assunto principal: Autoanálise / Vácuo / Sedimentação Sanguínea / Idoso / Feminino / Humanos / Masculino / Modelos Lineares / Criança / Pré-Escolar Tipo de estudo: Estudos de avaliação / Estudo prognóstico Idioma: Inglês Revista: Indian J Pathol Microbiol Ano de publicação: 2007 Tipo de documento: Artigo