Unidirectional valve malfunction by the breakage or malposition of disc: two cases report / 대한마취과학회지
Korean Journal of Anesthesiology
;
: 337-340, 2013.
Artigo
em Inglês
| WPRIM
| ID: wpr-100098
ABSTRACT
Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.