Your browser doesn't support javascript.
loading
Unidirectional valve malfunction by the breakage or malposition of disc: two cases report / 대한마취과학회지
Korean Journal of Anesthesiology ; : 337-340, 2013.
Artigo em Inglês | WPRIM | ID: wpr-100098
ABSTRACT
Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Respiração / Barotrauma / Ventilação / Capnografia / Hipercapnia Idioma: Inglês Revista: Korean Journal of Anesthesiology Ano de publicação: 2013 Tipo de documento: Artigo