Your browser doesn't support javascript.
loading
Investigation and analysis of energy release of cardiac defibrillator at different stages of lifecycle of that / 中国医学装备
China Medical Equipment ; (12): 39-43, 2024.
Article em Zh | WPRIM | ID: wpr-1026521
Biblioteca responsável: WPRO
ABSTRACT
Objective:To analyze the performance characteristics of the energy output of defibrillation device at different lifecycle stages of the equipment,and to improve the level of management,so as to ensure the safety and effectiveness of using the device.Methods:A total of 90 defibrillation devices of using 10 types included 861290 and CardioServ(included scrapped devices)during the period of 2015-2022 were retrospectively analyzed.The detected data of energy output of defibrillation device were analyzed as statistical method,and the error of releasing energy was calculated.A total of 36 defibrillation devices that were in the early stages of use(at the first three years of device use)were divided into the first year,the second year and the third year,and the data of energy outputs of devices among three years were compared.The data of the type A and type B defibrillation devices,which were the largest number of devices in the normal stage of use(the middle stage of use),were calibrated according to the energy release in the three years between 2018 and 2000.The difference of releasing energy at the preseted value of 100J between the two types of devices was analyzed.Finally,the errors of energy releases of 8 devices,which energy outputs exceeded the deadline,in the scrap period between 2015 and 2022 were summarized.Results:In the data of three groups of devices in the early stages of use,the differences at the first and second year of device use among 100J,150J and 200J of the energy releases of the preseted values were significant(t=-0.17,-0.17,-0.58,P>0.05).The difference of the measured values between the first and third years of device use was not significant(Z=-0.70,-0.38,-0.86,P>0.05).The results of variance analysis of repeated measurement of the energy releases of the devices in normal stage indicated that the difference of the energy release at 100J preseted point among different types of 41 devices was significant(F=4.40,P<0.05).The energy release of type X defibrillator appeared constantly high,and the relative error increased with the increasing of preseted values.The repeatability of the device was better,and the relationship between preseted energy(x)and release energy(y)conformed to linear relationship(R2=0.9985).In these defibrillation devices that were using,the qualified rate of energy output of>100J preseted point was 97.68%.Conclusion:There is slight difference in the mean value of energy release between different type of defibrillation devices within the qualified range,and the energy release still is a performance indicator that should be highly focused for defibrillation devices.We should combine with the maintenance and repair data of device to conduct in-depth analysis,so as to grasp the operating status of the device,and optimize the strategy of quality control,and ensure the safety of defibrillator in clinical use.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: China Medical Equipment Ano de publicação: 2024 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: China Medical Equipment Ano de publicação: 2024 Tipo de documento: Article