Korean Type Distal Radius Anatomical Volar Plate System: A Preliminary Report
Clinics in Orthopedic Surgery ; : 258-266, 2014.
Artigo em Inglês | WPRIM | ID: wpr-104732
ABSTRACT

BACKGROUND:

Distal radius fracture is the most common fracture of the upper extremity, and approximately 60,000 distal radius fractures occur annually in Korea. Internal fixation with an anatomical volar locking plate is widely used in the treatment of unstable distal radius fractures. However, most of the currently used distal radius anatomical plate systems were designed based on the anatomical characteristics of Western populations. Recently, the Korean-type distal radius anatomical volar plate (K-DRAVP) system was designed and developed based on the anatomical characteristics of the distal radius of Koreans. The purpose of this study was to evaluate the preliminary results of the new K-DRAVP system, and to compare its radiologic and functional results with those of the other systems.

METHODS:

From March 2012 to October 2012, 46 patients with acute distal radius fractures who were treated with the K-DRAVP system at three hospitals were enrolled in this study. Standard posteroanterior and lateral radiographs were obtained to assess fracture healing, and three radiographic parameters (volar tilt, radial inclination, and radial length) were assessed to evaluate radiographic outcomes. The range of motion and grip strength, the Gartland and Werley scoring system, and the disabilities of the arm, shoulder and hand (DASH) questionnaire were used to assess clinical and functional outcomes.

RESULTS:

All radiologic parameters were restored to normal values, and maintained without any loosening or collapse until the time of final follow-up. Grip strength was restored to 84% of the value for the unaffected side. The mean range of motion of the wrist at final follow-up was restored to 77%-95% of the value for the unaffected side. According to the Gartland and Werley scoring system, there were 16 excellent, 26 good, and 4 fair results. The mean DASH score was 8.4 points. There were no complications after surgery.

CONCLUSIONS:

The newly developed K-DRAVP system could be used to restore and maintain good anatomical parameters, and provide good clinical outcomes with low complication rates. This system is a promising surgical option for the treatment of distal radius fractures in the Korean population.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Rádio (Anatomia) / Fraturas do Rádio / Placas Ósseas / Amplitude de Movimento Articular / Placa Palmar / República da Coreia / Fixação Interna de Fraturas Limite: Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino País/Região como assunto: Ásia Idioma: Inglês Revista: Clinics in Orthopedic Surgery Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Rádio (Anatomia) / Fraturas do Rádio / Placas Ósseas / Amplitude de Movimento Articular / Placa Palmar / República da Coreia / Fixação Interna de Fraturas Limite: Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino País/Região como assunto: Ásia Idioma: Inglês Revista: Clinics in Orthopedic Surgery Ano de publicação: 2014 Tipo de documento: Artigo