Your browser doesn't support javascript.
loading
Digital Tomosynthesis for PNS Evaluation: Comparisons of Patient Exposure and Image Quality with Plain Radiography
Korean Journal of Radiology ; : 136-143, 2012.
Artigo em Inglês | WPRIM | ID: wpr-112476
ABSTRACT

OBJECTIVE:

We investigated low dose digital tomosynthesis (DT) for the evaluation of the paranasal sinus (PNS), and compared its diagnostic accuracy with a PNS radiography series (XR). MATERIALS AND

METHODS:

We enrolled 43 patients for whom XR, PNS DT, and OMU CT were performed. We measured effective doses (EDs) of XR, DT, and OMU CT using Monte Carlo simulation software. Two radiologists performed independent observation of both XR and DT. For seven PNSs, they scored anatomic conspicuity of sinuses and confidence on the presence of sinusitis using nine point scales. OMU CT was observed by the third radiologist and the findings were regarded as reference standard. We compared scores for conspicuity and sinusitis confidence between XR and DT.

RESULTS:

Mean EDs were 29 +/- 6 microSv, 48 +/- 10 microSv, and 980 +/- 250 microSv, respectively, for XR, DT, and CT. Mean scores for conspicuity were 6.3 and 7.4, respectively, for XR and DT. Sensitivity per patient basis for sinusitis detection were 52% and 96%, respectively, for XR and DT in observer 1 (p = 0.001) and 80% and 92% for observer 2 (p = 0.25). Specificities for sinusitis exclusion were 100% for both XR and DT for observer 1 and 89% and 100% for observer 2 (p = 0.50). Accuracies for sinusitis diagnosis were 72% and 98%, respectively, for XR and DT for observer 1 (p = 0.001) and 84% and 95% for observer 2 (p = 0.125).

CONCLUSION:

Patient radiation dose from low dose DT is comparable with that of PNS XR. Diagnostic sensitivity of DT for sinusitis was superior to PNS XR.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Seios Paranasais / Doses de Radiação / Sinusite / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Método de Monte Carlo / Curva ROC / Sensibilidade e Especificidade / Estatísticas não Paramétricas Tipo de estudo: Estudo diagnóstico / Avaliação Econômica em Saúde / Estudo prognóstico Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2012 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Seios Paranasais / Doses de Radiação / Sinusite / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Método de Monte Carlo / Curva ROC / Sensibilidade e Especificidade / Estatísticas não Paramétricas Tipo de estudo: Estudo diagnóstico / Avaliação Econômica em Saúde / Estudo prognóstico Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2012 Tipo de documento: Artigo