Guidewire Breakage during Neurointerventional Procedures: a Report of Two Cases
Korean Journal of Radiology
;
: 638-640, 2011.
Artigo
em Inglês
| WPRIM
| ID: wpr-116554
ABSTRACT
We report on two cases of microguidewire breakage that occurred during endovascular treatment of intracranial aneurysms. The microguidewire can be broken when a part of the wire is stuck due to vascular tortuosity, and, subsequently, application of excessive rotational movement. The mechanical and physical properties of a microguidewire are also important factors in microguidewire breakage. We also suggest technical tips for avoidance of this problem.