Influence of B₁-Inhomogeneity on Pharmacokinetic Modeling of Dynamic Contrast-Enhanced MRI: A Simulation Study
Korean Journal of Radiology ; : 585-596, 2017.
Artigo em Inglês | WPRIM | ID: wpr-118264
ABSTRACT

OBJECTIVE:

To simulate the B₁-inhomogeneity-induced variation of pharmacokinetic parameters on dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI). MATERIALS AND

METHODS:

B₁-inhomogeneity-induced flip angle (FA) variation was estimated in a phantom study. Monte Carlo simulation was performed to assess the FA-deviation-induced measurement error of the pre-contrast R₁, contrast-enhancement ratio, Gd-concentration, and two-compartment pharmacokinetic parameters (K(trans), v(e), and v(p)).

RESULTS:

B₁-inhomogeneity resulted in −23–5% fluctuations (95% confidence interval [CI] of % error) of FA. The 95% CIs of FA-dependent % errors in the gray matter and blood were as follows −16.7–61.8% and −16.7–61.8% for the pre-contrast R₁, −1.0–0.3% and −5.2–1.3% for the contrast-enhancement ratio, and −14.2–58.1% and −14.1–57.8% for the Gd-concentration, respectively. These resulted in −43.1–48.4% error for K(trans), −32.3–48.6% error for the v(e), and −43.2–48.6% error for v(p). The pre-contrast R₁ was more vulnerable to FA error than the contrast-enhancement ratio, and was therefore a significant cause of the Gd-concentration error. For example, a −10% FA error led to a 23.6% deviation in the pre-contrast R₁, −0.4% in the contrast-enhancement ratio, and 23.6% in the Gd-concentration. In a simulated condition with a 3% FA error in a target lesion and a −10% FA error in a feeding vessel, the % errors of the pharmacokinetic parameters were −23.7% for K(trans), −23.7% for v(e), and −23.7% for v(p).

CONCLUSION:

Even a small degree of B₁-inhomogeneity can cause a significant error in the measurement of pharmacokinetic parameters on DCE-MRI, while the vulnerability of the pre-contrast R₁ calculations to FA deviations is a significant cause of the miscalculation.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Encéfalo / Imageamento por Ressonância Magnética / Método de Monte Carlo / Imagens de Fantasmas / Substância Cinzenta Tipo de estudo: Avaliação Econômica em Saúde Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Encéfalo / Imageamento por Ressonância Magnética / Método de Monte Carlo / Imagens de Fantasmas / Substância Cinzenta Tipo de estudo: Avaliação Econômica em Saúde Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2017 Tipo de documento: Artigo