Reduction of the HIV Seroconversion Window Period and False Positive Rate by Using ADVIA Centaur HIV Antigen/Antibody Combo Assay
Annals of Laboratory Medicine
;
: 420-425, 2013.
Artigo
em Inglês
| WPRIM
| ID: wpr-118366
ABSTRACT
BACKGROUND:
Early diagnosis of HIV infection reduces morbidity and mortality. Fourth-generation HIV detection assays are more sensitive because they can detect p24 antigen as well as anti-HIV antibodies. In this study, we evaluated the performance of a new fourth-generation ADVIA Centaur HIV antigen/antibody combo (CHIV) assay (Siemens Healthcare Diagnostics Inc., USA) for early detection of HIV infection and reduction of false positive rate.METHODS:
Four seroconversion panels were included. The third-generation ADVIA Centaur HIV 1/O/2 enhanced (EHIV) assay (Siemens Healthcare Diagnostics Inc., USA) and fourth-generation CHIV assay were used to test each panel for HIV infection. The presence of antigen was confirmed using HIV p24 antigen assay. To evaluate false-positivity and specificity, 54 HIV false-positive and HIV-negative serum samples from 100 hospitalized patients and 600 healthy subjects were included.RESULTS:
Compared to the EHIV assay, the CHIV assay had a shorter window for three of the seroconversion panels a difference of 10 days and two bleeds in one panel, and 4 days and one bleed in the other two panels. Only 34 of the 54 (63%) samples known to yield false-positive results by EHIV assay had repeatedly yielded reactive results in the CHIV assay. One of the 600 healthy subjects had a false-positive result with the CHIV assay; thus, the specificity was 99.85% (699/700). CHIV accurately determined the reactive results for the HIV-confirmed serum samples from known HIV patients and Korea Food & Drug Administration (KFDA) panels.CONCLUSIONS:
The new fourth-generation ADVIA Centaur HIV assay is a sensitive and specific assay that shortens the serological window period and allows early diagnosis of HIV infection.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Kit de Reagentes para Diagnóstico
/
Fatores de Tempo
/
Anticorpos Anti-HIV
/
Sensibilidade e Especificidade
/
Soropositividade para HIV
/
Proteína do Núcleo p24 do HIV
/
Reações Falso-Positivas
/
República da Coreia
Tipo de estudo:
Estudo diagnóstico
/
Estudo de rastreamento
Limite:
Feminino
/
Humanos
/
Masculino
/
Gravidez
País/Região como assunto:
Ásia
Idioma:
Inglês
Revista:
Annals of Laboratory Medicine
Ano de publicação:
2013
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS