Point of Care D-Dimer Testing in the Emergency Department: A Bioequivalence Study
Annals of Laboratory Medicine ; : 34-38, 2013.
Artigo em Inglês | WPRIM | ID: wpr-119343
ABSTRACT

BACKGROUND:

D-dimer is used widely as a diagnostic aid in low- and moderate-risk patients with suspected venous thromboembolism (VTE). While our laboratory utilizes VIDAS D-dimer analyzer (bioMerieux SA, France), our emergency department (ED) recently procured a D-dimer analyzer AQT90 FLEX (Radiometer Medical ApS, Denmark) for point of care testing (POCT) to facilitate patient management. We aimed to determine whether the time taken to receive D-dimer results using the 2 different analyzers differed significantly and to quantify the limits of agreement between the results of the 2 methods measured on the same patient.

METHODS:

Adult patients presenting to the ED and requiring diagnostic workup for suspected VTE were included in this prospective observational study. Patients underwent simultaneous D-dimer measurements using the 2 different analyzers.

RESULTS:

The paired results from 104 patients were analyzed. The median time for the D-dimer results from triage by VIDAS was 258 min (Inter-quartile range [IQR], 173-360) and by POCT was 146 min (IQR, 55-280.5); the median time difference was 101.5 min (IQR, 82-125.5). On an average, POCT D-dimer values were 15% lower on the same sample (limits of agreement, 34-213%). POCT predicted 83% of VIDAS positive results (sensitivity, 83.3% [95% confidence interval (CI), 70.4-91.3%]; specificity, 100% [95% CI, 93.6-100%]). All patients with positive imaging were identified correctly by both methods.

CONCLUSIONS:

POCT delivers D-dimer results in significantly shorter turnaround times than pathology services; however, poor bioequivalence between VIDAS and POCT raises the issue of acceptability for use in the ED.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Kit de Reagentes para Diagnóstico / Fatores de Tempo / Produtos de Degradação da Fibrina e do Fibrinogênio / Tomografia Computadorizada por Raios X / Estudos Prospectivos / Sensibilidade e Especificidade / Sistemas Automatizados de Assistência Junto ao Leito / Serviço Hospitalar de Emergência / Tromboembolia Venosa Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo prognóstico Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2013 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Kit de Reagentes para Diagnóstico / Fatores de Tempo / Produtos de Degradação da Fibrina e do Fibrinogênio / Tomografia Computadorizada por Raios X / Estudos Prospectivos / Sensibilidade e Especificidade / Sistemas Automatizados de Assistência Junto ao Leito / Serviço Hospitalar de Emergência / Tromboembolia Venosa Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo prognóstico Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2013 Tipo de documento: Artigo