Your browser doesn't support javascript.
loading
An Unexpected Vapor Leakage from Unlocked Vaporizer (Penlon Sigma Delta Anesthetic Vaporizer): A case report / 대한마취과학회지
Korean Journal of Anesthesiology ; : 230-232, 2006.
Artigo em Coreano | WPRIM | ID: wpr-119951
ABSTRACT
We present a case of an anesthetic gas leak through the valve of the Selectatec back bar during a surgical procedure. The Selectatec Vaporizing System is a quick-change system consisting of anesthetic vaporizers of the 'Tec models' of the Ohmeda machine that are seated on a compatibility manifold block located on the back bar of the anesthetic machine. In order to prevent a leak and a failure of vapor delivery, the "Tec models series" can only be turned 'on' when the locking lever is turned to the 'lock' position. However in this case, the control knob was rotated despite the back bar locking mechanism not being engaged, and a gas leak developed through the unlocked valves in Selectatec back bar. The Penlon Sigma Delta Anesthetic Vaporizer (sevoflurane) in a Datex-Ohmeda Aestiva/5 anesthesia machine can be turned 'on' when the locking lever is maintained at the 'unlock' position.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Nebulizadores e Vaporizadores / Anestesia Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2006 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Nebulizadores e Vaporizadores / Anestesia Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2006 Tipo de documento: Artigo