Normal Value of Standard Electroretinography and Change with Age and Sex(II)-Results Using Burian-Allen Electrode and Comparison with Results Using EFG-jet Electrode-
Journal of the Korean Ophthalmological Society
;
: 128-139, 1999.
Artigo
em Coreano
| WPRIM
| ID: wpr-120421
ABSTRACT
Using Burian-Allen bipolar contact lens electrode, Authors examined the normal 120 eyes of 120 subjects(male 57, female 63), age ranged 7 to 83 years(mean 44.5 years), with the method which was same as the ISCEV standard. There was no statistically significant differences between the value obtained using Burian-Allen electrode and the value obtained using ERG-jet electrode(p>0.05). We expect that this data would be useful testing method to evaluate the retinal disease. And this data may provide new reference value in the study of electroretinogram.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Valores de Referência
/
Doenças Retinianas
/
Eletrodos
/
Eletrorretinografia
Limite:
Feminino
/
Humanos
Idioma:
Coreano
Revista:
Journal of the Korean Ophthalmological Society
Ano de publicação:
1999
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS