Flow Cytometric Human Leukocyte Antigen-B27 Typing with Stored Samples for Batch Testing
Annals of Laboratory Medicine ; : 174-183, 2013.
Artigo em Inglês | WPRIM | ID: wpr-144110
ABSTRACT

BACKGROUND:

Flow cytometry (FC) HLA-B27 typing is still used extensively for the diagnosis of spondyloarthropathies. If patient blood samples are stored for a prolonged duration, this testing can be performed in a batch manner, and in-house cellular controls could easily be procured. In this study, we investigated various methods of storing patient blood samples.

METHODS:

We compared four storage

methods:

three methods of analyzing lymphocytes (whole blood stored at room temperature, frozen mononuclear cells, and frozen white blood cells [WBCs] after lysing red blood cells [RBCs]), and one method using frozen platelets (FPLT). We used three ratios associated with mean fluorescence intensities (MFI) for HLAB27 assignment the B27 MFI ratio (sample/control) for HLA-B27 fluorescein-5-isothiocyanate (FITC); the B7 MFI ratio for HLA-B7 phycoerythrin (PE); and the ratio of these two ratios, B7/B27 ratio.

RESULTS:

Comparing the B27 MFI ratios of each storage method for the HLA-B27+ samples and the B7/B27 ratios for the HLA-B7+ samples revealed that FPLT was the best of the four methods. FPLT had a sensitivity of 100% and a specificity of 99.3% for HLA-B27 assignment in DNA-typed samples (N=164) when the two criteria, namely, B27 MFI ratio >4.0 and B7/B27 ratio <1.5, were used.

CONCLUSIONS:

The FPLT method was found to offer a simple, economical, and accurate method of FC HLA-B27 typing by using stored patient samples. If stored samples are used, this method has the potential to replace the standard FC typing method when used in combination with a complementary DNA-based method.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Temperatura / Plaquetas / Teste de Histocompatibilidade / Leucócitos Mononucleares / Antígeno HLA-B7 / Antígeno HLA-B27 / Espondiloartropatias / Eritrócitos / Reação em Cadeia da Polimerase em Tempo Real / Citometria de Fluxo Limite: Humanos Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2013 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Temperatura / Plaquetas / Teste de Histocompatibilidade / Leucócitos Mononucleares / Antígeno HLA-B7 / Antígeno HLA-B27 / Espondiloartropatias / Eritrócitos / Reação em Cadeia da Polimerase em Tempo Real / Citometria de Fluxo Limite: Humanos Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2013 Tipo de documento: Artigo