ABSTRACT

BACKGROUND:

The importance and usefulness of therapeutic drug monitoring (TDM) have been emphasized, and analysis of drugs has been increased in clinical laboratories. We evaluated the analytical performance and clinical usefulness of a recently introduced enzyme multiplied immunoassay instrument, Viva-E Drug Testing System (Dade Behring Inc., USA).

METHODS:

Using patients' samples and quality control material, we evaluated the analytical performance of Viva-E for a total of 11 drugs (cyclosporine, tacrolimus, mycophenolic acid, valproic acid, digoxin, theophylline, carbamazepine, phenytoin, phenobarbital, vancomycin, and gentamicin) with respect to linearity, precision, and correlations with other methods according to CLSI guidelines. Cobas Integra 800 (Roche Diagnostics, Switzerland) and API 4000 LC-MS/MS System (Applied Biosystems, USA) were used to make a comparison. In addition, we analyzed analysis time.

RESULTS:

Viva-E showed a good linearity (r2 > or = 0.97) for all items. Within-run CVs were within 5% and total CVs were within 10% for all drugs except for tacrolimus and digoxin at low concentrations. The system correlated well with the other methods (r=0.9283-0.9778). The time required for reporting the first sample was 11 min and the analysis time was 1.1 min.

CONCLUSIONS:

Since Viva-E showed a good analytical performance required for TDM in its linearity, precision, and accuracy with its wide drug menus including cyclosporine, tacrolimus, and mycophenolic acid, stat and random accessing functions, and the consolidation to a single workstation, it could be very useful in the clinical laboratory for various needs.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Padrões de Referência / Preparações Farmacêuticas / Reprodutibilidade dos Testes / Interpretação Estatística de Dados / Técnicas Imunoenzimáticas / Monitoramento de Medicamentos / Técnica de Imunoensaio Enzimático de Multiplicação Tipo de estudo: Guia de Prática Clínica Limite: Humanos Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Padrões de Referência / Preparações Farmacêuticas / Reprodutibilidade dos Testes / Interpretação Estatística de Dados / Técnicas Imunoenzimáticas / Monitoramento de Medicamentos / Técnica de Imunoensaio Enzimático de Multiplicação Tipo de estudo: Guia de Prática Clínica Limite: Humanos Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2007 Tipo de documento: Artigo