Comparison of Nine Different Qualitative HBsAg Assay Kits / 대한진단검사의학회지
The Korean Journal of Laboratory Medicine ; : 178-184, 2010.
Artigo em Coreano | WPRIM | ID: wpr-151622
ABSTRACT

BACKGROUND:

Qualitative hepatitis B surface antigen (HBsAg) assay kits are still commonly used in Korea where hepatitis B virus (HBV) infection is endemic. The accurate determination of HBsAg plays a crucial role in the diagnosis and prevention of HBV infection, especially in endemic areas. The aim of this study was to compare the detection sensitivities of 9 qualitative HBsAg assay kits.

METHODS:

Seven pooled sera with HBsAg concentration ranging from 0.14 IU/mL to 29.96 IU/mL were prepared. The HBsAg concentration of each pooled serum was determined by a quantitative HBsAg assay, Architect HBsAg (Abbott Laboratories, Ireland). The fully automated immunoassay kits included Elecsys HBsAg (Roche Diagnostics, Germany) and Immulite 2000 HBsAg (DPC, USA) and the rapid tests included 5 immunochromatographic assay (ICA) kits and 2 reverse passive hemagglutination assay (RPHA) kits.

RESULTS:

Elecsys HBsAg (Roche Diagnostics) showed positive result in pooled serum with HBsAg concentration of 0.14 IU/mL, but Immulite 2000 HBsAg (DPC) showed negative result in the same concentration. Although ICA kits showed variable results among different assay kits, all of them showed negative results in pooled sera with HBsAg concentration of < or =1.89 IU/mL. Two RPHA kits showed negative results in pooled sera with HBsAg concentration of < or =7.98 IU/mL.

CONCLUSIONS:

Although ICAs were more sensitive than RPHAs, they had variable sensitivities for HBsAg and were less sensitive than the automated immunoassay kits. Therefore, ICAs and RPHAs should be used with caution in the screening tests for HBsAg and their sensitivities need to be improved.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Kit de Reagentes para Diagnóstico / Ensaio de Imunoadsorção Enzimática / Testes de Hemaglutinação / Técnicas Eletroquímicas / Genótipo / Hepatite B / Antígenos de Superfície da Hepatite B / Medições Luminescentes Tipo de estudo: Pesquisa qualitativa Limite: Humanos Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2010 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Kit de Reagentes para Diagnóstico / Ensaio de Imunoadsorção Enzimática / Testes de Hemaglutinação / Técnicas Eletroquímicas / Genótipo / Hepatite B / Antígenos de Superfície da Hepatite B / Medições Luminescentes Tipo de estudo: Pesquisa qualitativa Limite: Humanos Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2010 Tipo de documento: Artigo