A Prospective Study on the lncidence of Ventilator-associated Pneumonia in Patients with Circuit Changes every 3 days Versus Weekly Changes / 병원감염관리
Korean Journal of Nosocomial Infection Control ; : 9-21, 2000.
Artigo em Coreano | WPRIM | ID: wpr-151746
ABSTRACT

BACKGROUND:

Ventilator associated pneumonia (VAP) is the most serious nosocomial infection of intensive care units. Several studies have investigated the relationship between the interval of ventilator circuit changes and the incidence of pneumonia in foreign countries, but there are no reports about it in Korea yet. So we performed this study to compare the clinical and cost impact between 3 days and 7 days interval in ventilator circuit changes.

METHODS:

Seoul National University Hospital is a 1,500-bed, university affiliated, tertiary and acute care hospital. All patients admitted to medical intensive care unit (MICU) and surgical intensive care unlt (SICU) between April 1, 1998 and October 31, 1998, requiring mechanical ventilation were included. Patients were divided into two groups of a-cay circuit changes and weekly changes. Daily surveillance was conducted using the criteria of VAP of the National Nosocomial Infection Surveillance System. Incidence of VAP and risk factors for VAP were evaluated. Standard microbiologic methods were used for the identification of clinical and environmental isolates. Statistical analysis was done by SAS Program (version 6.12), analysis of difference in variables was performed using chi-square test and t-test. Analysis of odds ratios was done with logistic regression analysis.

RESULTS:

VAP developed at a rate of 12.2 per 1,000 ventilator-days in the 3 days change group and 15.6 per 1,000 ventilator-days in the weekly change group (P=0.7240). The only statistically significant risk factor of VAP was duration of mechanical ventilation, The risk of VAP in patients with more than 7 days was 2.23 times higher than in patients with 7 days and below (OR; 2.2296). Estimated annual savings of nursing time by extending ventilator circuit change interval from 3 days to 7 days were 26,806 min 48 sec and estimated savings of cost by reduction of nursing times was calculated as 6,701,700 won.

CONCLUSIONS:

Weekly ventilator circuit changes in patients undergoing ventilation therapy in the ICU do not contribute to increased the rates of VAP and are cost-effective.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Pneumonia / Respiração Artificial / Ventilação / Ventiladores Mecânicos / Modelos Logísticos / Razão de Chances / Infecção Hospitalar / Incidência / Estudos Prospectivos / Fatores de Risco Tipo de estudo: Estudo de etiologia / Estudo de incidência / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Humanos País/Região como assunto: Ásia Idioma: Coreano Revista: Korean Journal of Nosocomial Infection Control Ano de publicação: 2000 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Pneumonia / Respiração Artificial / Ventilação / Ventiladores Mecânicos / Modelos Logísticos / Razão de Chances / Infecção Hospitalar / Incidência / Estudos Prospectivos / Fatores de Risco Tipo de estudo: Estudo de etiologia / Estudo de incidência / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Humanos País/Região como assunto: Ásia Idioma: Coreano Revista: Korean Journal of Nosocomial Infection Control Ano de publicação: 2000 Tipo de documento: Artigo