ABSTRACT
We experienced a case of ventilation impairment resulted by the wrong-assembled unidirectional valves passed the automatic checkout of ventilator during general anesthesia. A 63-year-old man was scheduled for Ivor-Lewis operation due to esophageal cancer. After endotracheal intubation, ventilation impairment and abdominal distension were developed. No abnormal finding was found with bronchofiberoscopic examination. Suspecting anesthetic equipment failure, unidirectional valves were separated and they were proved to be connected upside-down.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Ventilação / Neoplasias Esofágicas / Ventiladores Mecânicos / Falha de Equipamento / Intubação Intratraqueal / Anestesia Geral Limite: Humanos Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2008 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Ventilação / Neoplasias Esofágicas / Ventiladores Mecânicos / Falha de Equipamento / Intubação Intratraqueal / Anestesia Geral Limite: Humanos Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2008 Tipo de documento: Artigo