Evaluation of a Chest Circumference-Adapted Protocol for Low-Dose 128-Slice Coronary CT Angiography with Prospective Electrocardiogram Triggering
Korean Journal of Radiology ; : 13-20, 2015.
Artigo em Inglês | WPRIM | ID: wpr-157431
ABSTRACT

OBJECTIVE:

To assess the effect of chest circumference-adapted scanning protocol on radiation exposure and image quality in patients undergoing prospective electrocardiogram (ECG)-triggered coronary CT angiography (CCTA). MATERIALS AND

METHODS:

One hundred-eighty-five consecutive patients, who had undergone prospective ECG triggering CCTA with a 128-slice CT, were included in the present study. Nipple-level chest circumference, body weight and height were measured before CT examinations. Patients were divided into four groups based on kV/ref.mAs = 100/200, 100/250, 120/200, and 120/250, when patient's chest circumference was 95.0 (n = 32), respectively. Image quality per-segment was independently assessed by two experienced observers. Image noise and attenuation were also measured. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The effective radiation dose was calculated using CT dose volume index and the dose-length product.

RESULTS:

A significant correlation was observed between patients' chest circumference and body mass index (r = 0.762, p < 0.001). Chest circumference ranged from 74 to 105 cm, and the mean effective radiation dose was 1.9-3.8 mSv. Diagnostic image quality was obtained in 98.5% (2440/2478) of all evaluated coronary segments without any significant differences among the four groups (p = 0.650). No significant difference in image noise was observed among the four groups (p = 0.439), thus supporting the validity of the chest circumference-adapted scanning protocol. However, vessel attenuation, SNR and CNR were significantly higher in the 100 kV groups than in the 120 kV groups (p < 0.05).

CONCLUSION:

A measure of chest circumference can be used to adapt tube voltage and current for individualized radiation dose control, with resultant similar image noise and sustained diagnostic image quality.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Doses de Radiação / Tomografia Computadorizada por Raios X / Índice de Massa Corporal / Estudos Prospectivos / Meios de Contraste / Eletrocardiografia / Circunferência da Cintura / Razão Sinal-Ruído / Coração / Frequência Cardíaca Tipo de estudo: Guia de Prática Clínica / Estudo observacional Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Doses de Radiação / Tomografia Computadorizada por Raios X / Índice de Massa Corporal / Estudos Prospectivos / Meios de Contraste / Eletrocardiografia / Circunferência da Cintura / Razão Sinal-Ruído / Coração / Frequência Cardíaca Tipo de estudo: Guia de Prática Clínica / Estudo observacional Limite: Adulto / Idoso / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2015 Tipo de documento: Artigo