A Study of Drug Content and Cell Cytotoxicity of Paclitaxel-eluting Stents Coated with Various Biopolymer / 의학물리
Korean Journal of Medical Physics
;
: 125-131, 2009.
Artigo
em Coreano
| WPRIM
| ID: wpr-158395
ABSTRACT
In this study, the paclitaxel eluting stent (PES) was prepared by coating a biliary stent with paclitaxel using various biopolymer such as poly (vinyl acetate) (PVAc), poly (lactic-co-glycolic acid) (PLGA), Silicone rubber for restenosis prevention in gastrointestinal disease by a dip-coating method. Drug contents of PES were increased as surface area of stent, concentration and molecular weight of coating polymer increase. In 1H-NMR specta, we know that drug did not change by confirming specific peaks of paclitaxel in PES. As shown in SEM image, PES prepared using various biopolymer is coated clearly and regularly except Silicone rubber coating polymer. In in vitro cell cytotoxicity test, bare stent showed low cytotoxic effect against CT-26 colon carcinoma cell line on 3 day. However, PES coated with PLGA 502H showed the highest cytotoxicity because PLGA 502H is biodegradable polymer and has less molecular weight than other coating polymer. These results suggest that PES coated various biopolymer can be prevented restenosis in gastrointestinal disease.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Ácido Poliglicólico
/
Polímeros
/
Elastômeros de Silicone
/
Biopolímeros
/
Stents
/
Linhagem Celular
/
Paclitaxel
/
Colo
/
Ácido Láctico
/
Gastroenteropatias
Idioma:
Coreano
Revista:
Korean Journal of Medical Physics
Ano de publicação:
2009
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS