Your browser doesn't support javascript.
loading
Standardization for Dual Energy X-ray Absorptiometry / 대한내과학회지
Korean Journal of Medicine ; : 445-449, 1997.
Artigo em Coreano | WPRIM | ID: wpr-160829
ABSTRACT

OBJECTIVES:

Bone densitometry is a current method for evaluating skeletal status, assessing osteoporosis, and determining fracture risk. DEXA has rapidly become a dominent method for evaluating skeletal status. But the comparison of patient data among different DEXA scanners is complicated because the instruments show differences in scanner design, bone mineral calibration, and analysis algorithms. The purpose of this study is making standardization of DEXA and comparability among different DEXA system.

METHODS:

Posteroanterior lumbar spine (L2-IA) measurements of healthy 83 racially homogenous Korean women, age 38-66 years (mean +/- 516, range of BMD 0.624-1.574g/cm) were obtained on a Lunar DPX-L, a Hologic QDR-2000, and a Norland XR-36. All sujects had no spinal deformities, ostophytes, fracture and scoliosis on thoracolumbar spinal X-ray. We performed the measurement of BMD in each subjects with different three scanners at the same time.

RESULTS:

The results of cross-calibration spinal BMD (L2-4, g/cm2) in patient study (r2=0.972-0.974); Lunar DPX-L=(1.149 X Hologic)+0.008=(1.022 X Norland)+0.137, Hologic QDR-2000=(0.847 X Lunar)+0.019= (0.876 X Norland)+0.124, and Norland XR-36=(0.954 X Lunar)-0.107=1.110 X Hologic)-0.113. The average absolute difference in patients` spinal BMD values (L2-L4) between Lunar and Hologic was 0.146g/cm2 (14%); it was 0.156g/cm2 (15%) between Lunar and Norland and 0.010g/cm2 (1%) between Hologic and Norland.

CONCLUSIONS:

The formula of this cross-calibration shows good correlation. We conclude that it is possible to compare the different spinal BMD values obtained on the different DEXA scanners.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Osteoporose / Escoliose / Coluna Vertebral / Anormalidades Congênitas / Calibragem / Absorciometria de Fóton / Densidade Óssea / Densitometria Limite: Feminino / Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 1997 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Osteoporose / Escoliose / Coluna Vertebral / Anormalidades Congênitas / Calibragem / Absorciometria de Fóton / Densidade Óssea / Densitometria Limite: Feminino / Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 1997 Tipo de documento: Artigo