Your browser doesn't support javascript.
loading
New Leukocyte Counting Method of Cerebrospinal Fluid: Using Transparent Ruler Tape / 대한진단검사의학회지
The Korean Journal of Laboratory Medicine ; : 394-399, 2007.
Artigo em Coreano | WPRIM | ID: wpr-161978
ABSTRACT

BACKGROUND:

To enumerate leukocyte count in cerebrospinal fluid (CSF) is important for diagnosing bacterial meningitis. Using automated hematology analyzer for enumeration of leukocyte in CSF is below the sensitivity, so microscopic hemocytometric method is standard method. But this requires sufficient practical experience and has limitation of accuracy and stability. So we developed new manual method and evaluated it.

METHODS:

We designed new method using transparent ruler tape. We performed correlation, accuracy and precision test by counting leukocyte in diluted EDTA blood with three

methods:

new method, Neubauer and Nageotte hemocytometry. Twenty two CSF were used for stability test, which determines leukocyte count according to time (within one hour and after 2, 4 and 12 hr), by new method and Neubauer hemocytometry at room temperature.

RESULTS:

There was no clinical significant difference between three methods in correlation test, whereas Neubauer and Nageotte hemocytometry showed a bias to underestimation relative to the results obtained with new method in case with low leukocyte count. The new method showed the lowest CV and most accurate result. In stability test, leukocyte counts decreased being 44.4%, 72.1% of initial values after 2 hr, 14.8%, 31.1%, after 4 hr and 4.2%, 8.7%, after 12 hr, by Nageotte hemocytometry and new method, respectively.

CONCLUSIONS:

The new method we devised is simple, easy and applicable to use in a laboratory and offers advantages of improved precision and stability. It may be sufficient for replacing standard methods for leukocyte counting in CSF.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Fatores de Tempo / Líquido Cefalorraquidiano / Análise de Regressão / Reprodutibilidade dos Testes / Sensibilidade e Especificidade / Contagem de Leucócitos Tipo de estudo: Estudo diagnóstico / Guia de Prática Clínica Limite: Feminino / Humanos / Masculino Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Fatores de Tempo / Líquido Cefalorraquidiano / Análise de Regressão / Reprodutibilidade dos Testes / Sensibilidade e Especificidade / Contagem de Leucócitos Tipo de estudo: Estudo diagnóstico / Guia de Prática Clínica Limite: Feminino / Humanos / Masculino Idioma: Coreano Revista: The Korean Journal of Laboratory Medicine Ano de publicação: 2007 Tipo de documento: Artigo