Your browser doesn't support javascript.
loading
Percutaneous Retrieval and Redeployment of an Atrial Septal Occluder under Three-Dimensional Transesophageal Echocardiographic Guidance: A Case Report
Journal of Korean Medical Science ; : 871-873, 2014.
Artigo em Inglês | WPRIM | ID: wpr-163313
ABSTRACT
Percutaneous device closure for secundum atrial septal defects (ASDs) has been performed commonly and safely with high success rates. However, it is still challenging to close ASDs that are surrounded with deficient or hypermobile rims and could be compromised with an unexpected migration of device. We report a case of percutaneous Amplazter Septal Occluder (ASO; St. Jude Medical Inc., St. Paul, Minnesota, USA) device closure for an ASD with a thin and floppy interatrial septum, which immediately migrated into the right atrium and was not pulled back into the delivery sheath. To our knowledge, this is the first report on a successful percutaneous retrieval and redeployment of the device in such a situation, preventing any vascular injury or unplanned emergency open heart surgery.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Fluoroscopia / Ecocardiografia Transesofagiana / Ecocardiografia Tridimensional / Dispositivo para Oclusão Septal / Átrios do Coração / Comunicação Interatrial Tipo de estudo: Guia de Prática Clínica Limite: Feminino / Humanos Idioma: Inglês Revista: Journal of Korean Medical Science Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Fluoroscopia / Ecocardiografia Transesofagiana / Ecocardiografia Tridimensional / Dispositivo para Oclusão Septal / Átrios do Coração / Comunicação Interatrial Tipo de estudo: Guia de Prática Clínica Limite: Feminino / Humanos Idioma: Inglês Revista: Journal of Korean Medical Science Ano de publicação: 2014 Tipo de documento: Artigo