Usefulness of Computerized Objective Visual Acuity Test System Using Suppression Method
Journal of the Korean Ophthalmological Society
;
: 1127-1134, 2008.
Artigo
em Coreano
| WPRIM
| ID: wpr-164610
ABSTRACT
PURPOSE:
The objective visual acuity test is mandatory in certain cases, such as infants, nonverbal subjects, and subjects who need legal assistance. We designed an objective system for visual acuity test (SOVAT) consisting of three components stimuli applied via a suppression method, display and evaluation and made a suppression method as stimuli component for SOVAT. Usefulness of the SOVAT was evaluated.METHODS:
The visual stimuli were presented on a high-resolution head-mounted display (HMD). An eye movement tracking program and gaze monitoring device allowed us to monitor the patient's fixation status during the test. The suppression method, in addition to a conventional induction method, was developed to use with the SOVAT and its accuracy and confidence level were evaluated.RESULTS:
On the basis of clinical data, we present the reference values for the SOVAT as below. For the induction method, objective visual acuity (smallest pixel size), the presumed subjective visual acuity was 3 0.35-0.6, 7 0.05-0.25, 12 below 0.05 and for the suppression method it was 1 0.6-1.0, 1.5 0.4-0.7, 3 0.15-0.4, 5 0.1-0.2.CONCLUSIONS:
SOVAT using this suppression method to assess objective visual acuity proved to be more useful than that with a conventional induction method, especially in the range of high visual acuity.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Compostos Organotiofosforados
/
Valores de Referência
/
Atletismo
/
Acuidade Visual
/
Nistagmo Optocinético
/
Movimentos Oculares
Limite:
Humanos
/
Lactente
Idioma:
Coreano
Revista:
Journal of the Korean Ophthalmological Society
Ano de publicação:
2008
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS